TREO ABDOMINAL STENT-GRAFT SYSTEM
Report
- Report Number
- 2247858-2020-00035
- Event Type
- Malfunction
- Date Received
- July 13, 2020
- Date of Event
- June 19, 2020
- Report Date
- September 11, 2020
- Manufacturer
- BOLTON MEDICAL, INC.
- Product Code
- MIH
- PMA / PMN Number
- P190015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
"THE PATIENT HAD TORTUOUS ILIAC ACCESS. INITIALLY THE PHYSICIAN TRIED TO PLACE THE MAIN BODY UP THE RIGHT COMMON FEMORAL AND IT WOULD NOT PASS BEYOND A KNOWN CALCIFIC STENOSIS IN THE RIGHT COMMON ILIAC. THE PHYSICIAN THEN PUT THE DEVICE UP THE LEFT COMMON FEMORAL AND IT ADVANCED TO THE DESIRED LOCATION NEAR THE RENAL ARTERIES. HE THEN STARTED DEPLOYMENT OF THE GRAFT PER IFU. THE CONTRALATERAL GATE WAS UNCOVERED, AND THE PHYSICIAN PROCEEDED TO UNCLASP THE PROXIMAL CLASP. HE A LITTLE DIFFICULTY WITH THE GREY RELEASE RETAINER AND I REMINDED HIM AGAIN THAT IT WAS A PUSH IN THEN TWIST. WE WERE ABLE TO THE MOVE THE GREY RELEASE RETAINER AND PULL THE BLACK PROXIMAL CLASP HOLDER BACK. UNDER FLOURO THE CLASP MOVED BACK BUT IT APPEARED THAT THE SUPRARENAL STENT DID NOT DISENGAGE FROM THE APEX HOLDER. WE CONTINUED WITH THE PROCEDURE BY CANNULATING THE CONTRALATERAL GATE AND DEPLOYING THE CONTRALATERAL LIMB. THAT PROCESS WAS UNEVENTFUL. WE THEN PROCEEDED WITH THE DEPLOYMENT OF THE IPSILATERAL GATE. ONCE THE CATHETER WAS FREE FROM THE CONTRALATERAL GATE WE TRIED TO MOVE THE CATHETER AND NOTED THAT IT WAS STILL CONSTRAINED PROXIMALLY. THE PHYSICIAN TRIED TO MOVE THE CATHETER AND THE SUPRARENAL STENTS WERE MOVING. I HAD HIM GENTLY TWIST THE CATHETER AND YOU COULD SEE THE SUPRARENAL STENTS DISENGAGING, ONE BY ONE, FROM THE APEX HOLDER. WE THEN REMOVED THE DEVICE AND IMPLANTED THE IPSILATERAL LIMB." PATIENT OUTCOME: "(B)(6) SPOKE TO THE PHYSICIAN THE MORNING OF (B)(6) 2020 AND SHE IS DOING WELL ACCORDING TO DR. (B)(6). HE SAID SHE SHOULD BE RELEASED FROM THE HOSPITAL TOMORROW. "
THE TREO ABDOMINAL STENT GRAFT SYSTEM RECEIVED FDA APPROVAL FOR SALE IN THE US ON MAY 4, 2020 (P190015).
"THE PATIENT HAD TORTUOUS ILIAC ACCESS. INITIALLY THE PHYSICIAN TRIED TO PLACE THE MAIN BODY UP THE RIGHT COMMON FEMORAL AND IT WOULD NOT PASS BEYOND A KNOWN CALCIFIC STENOSIS IN THE RIGHT COMMON ILIAC. THE PHYSICIAN THEN PUT THE DEVICE UP THE LEFT COMMON FEMORAL AND IT ADVANCED TO THE DESIRED LOCATION NEAR THE RENAL ARTERIES. HE THEN STARTED DEPLOYMENT OF THE GRAFT PER IFU. THE CONTRALATERAL GATE WAS UNCOVERED, AND THE PHYSICIAN PROCEEDED TO UNCLASP THE PROXIMAL CLASP. HE A LITTLE DIFFICULTY WITH THE GREY RELEASE RETAINER AND I REMINDED HIM AGAIN THAT IT WAS A PUSH IN THEN TWIST. WE WERE ABLE TO THE MOVE THE GREY RELEASE RETAINER AND PULL THE BLACK PROXIMAL CLASP HOLDER BACK. UNDER FLOURO THE CLASP MOVED BACK BUT IT APPEARED THAT THE SUPRARENAL STENT DID NOT DISENGAGE FROM THE APEX HOLDER. WE CONTINUED WITH THE PROCEDURE BY CANNULATING THE CONTRALATERAL GATE AND DEPLOYING THE CONTRALATERAL LIMB. THAT PROCESS WAS UNEVENTFUL. WE THEN PROCEEDED WITH THE DEPLOYMENT OF THE IPSILATERAL GATE. ONCE THE CATHETER WAS FREE FROM THE CONTRALATERAL GATE WE TRIED TO MOVE THE CATHETER AND NOTED THAT IT WAS STILL CONSTRAINED PROXIMALLY. THE PHYSICIAN TRIED TO MOVE THE CATHETER AND THE SUPRARENAL STENTS WERE MOVING. I HAD HIM GENTLY TWIST THE CATHETER AND YOU COULD SEE THE SUPRARENAL STENTS DISENGAGING, ONE BY ONE, FROM THE APEX HOLDER. WE THEN REMOVED THE DEVICE AND IMPLANTED THE IPSILATERAL LIMB." PATIENT OUTCOME: "(B)(6) SPOKE TO THE PHYSICIAN THE MORNING OF (B)(6) 2020 AND SHE IS DOING WELL ACCORDING TO DR. FOTEH. HE SAID SHE SHOULD BE RELEASED FROM THE HOSPITAL TOMORROW. "
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 732926 | TREO ABDOMINAL STENT-GRAFT SYSTEM | STENT, ENDOVASCUALR GRAFT, AORTIC | MIH | BOLTON MEDICAL, INC. | 2006020088 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR |