PICCOLO XPRESS CHEMISTRY ANALYZER
Report
- Report Number
- 2939693-2020-00047
- Event Type
- Malfunction
- Date Received
- July 13, 2020
- Date of Event
- January 24, 2018
- Report Date
- June 25, 2021
- Manufacturer
- ABAXIS, INC.
- Product Code
- JJG
- PMA / PMN Number
- K934592
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
AS PART OF A THREE-YEAR RETROSPECTIVE REVIEW OF COMPLAINTS FOR THE QUALITY IMPROVEMENT PROCESS, CALLS WERE IDENTIFIED IN WHICH THE CUSTOMER EXPERIENCED BURNING ODOR AND/OR SMOKE. THERE WERE NO INJURIES RELATED TO THE CUSTOMER COMPLAINTS RECEIVED, BASED ON THE POTENTIAL OF BURNING SMELL/SMOKE TO LEAD TO A POTENTIAL INJURY. BASED ON MDR REGULATIONS, ABAXIS REPORTED THESE EVENTS DUE TO: THE LIKELIHOOD OF THIS MALFUNCTION TO LEAD TO AN ADVERSE EVENT, ALTHOUGH THERE WAS NO EVIDENCE OR REPORTS OF ASSOCIATED ADVERSE EVENTS. MINIMAL DETAILED DOCUMENTATION RELATED TO THE LIKELIHOOD AND CAUSE OF BURNING ODOR AND SMOKE. AS PART OF AN INVESTIGATION REGARDING THE OVERALL REPORTED MALFUNCTION OF A BURNING ODOR AND/OR SMOKING, RISK ASSESSMENT, AND MAUDE DATABASE SEARCH, ZOETIS HAS CONCLUDED THAT THE BURNING ODOR/SMOKE OVERHEATING COMPLAINTS ARE A LOW INJURY RISK FOR THE USER OR PATIENT. THE IDENTIFIED MALFUNCTIONS HAVE NOT CAUSED OR CONTRIBUTED TO A DEATH OR SERIOUS INJURY NOR DOES THE DATA SUGGEST THAT THE DEVICE OR A SIMILAR DEVICE MARKETED WOULD BE LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF THE MALFUNCTION WERE TO REOCCUR. ZOETIS WILL CONTINUE TO MONITOR COMPLAINTS FOR BURNING ODOR/SMOKING/OVERHEATING TO DETERMINE IF ANY NEW INFORMATION IS OBTAINED AND ANY CHANGE TO THE RISKS FOR PATIENTS OR USERS OF THE PICCOLO XPRESS ANALYZER. IF NEW INFORMATION IS OBTAINED, NEW CAUSES WHERE RISK HAS NOT BEEN EVALUATED, POTENTIAL OF SERIOUS INJURY REPORTED OR IDENTIFIED, THE SITE WILL FOLLOW THE REQUIRED PROCESS FOR MDR REPORTING AND TIMELINESS.
ON 24-JAN-2018, ABAXIS RECEIVED A CALL FROM THE CUSTOMER LAB REPORTING AN ISSUE WITH ANALYZER SN: (B)(4) STATING THAT THE DEVICE DRAWER WAS STUCK AND EMITTED SMOKE AND SPARKS. NO FIRE OR INJURY WAS REPORTED. THE CUSTOMER SENT THE DEVICE BACK FOR REPAIR. DURING EVALUATION OF THE DEVICE, THE REPORTED PROBLEM OF THE DRAWER JAMMING WAS CONFIRMED BY THE DEVICE LOGS. ROTOR ONE TESTING WAS RUN OVERNIGHT WITH NO ERRORS. NO SMOKING OR BURNING SMELL WAS NOTICED AND NO BURNED COMPONENTS OF THE DEVICE WERE FOUND. THE POWER PCB BOARD AND FAN FILTER WERE REPLACED AS A PRECAUTION. THE DEVICE PASSED ALL TESTING, WAS CLEANED, AND SENT BACK TO THE CUSTOMER SITE WHERE IT REMAINS. NO FURTHER ACTIONS WERE REQUIRED. THIS MDR IS BEING SUBMITTED AS A RESULT OF A THREE (3) YEAR RETROSPECTIVE REVIEW OF CLOSED COMPLAINTS ABAXIS PERFORMED AS PART OF ITS COMMITMENT TO CORRECT THE FDA-483 OBSERVATIONS RECEIVED ON AUGUST 22, 2019. THERE HAS BEEN NO REOCCURRENCE REGARDING THE ISSUE OF BURNING SMOKING WITH THIS DEVICE POST REPAIRS.
THE CUSTOMER LAB REPORTED THAT THE DEVICE DRAWER WAS STUCK, EMITTED SMOKE AND SPARKS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 729135 | PICCOLO XPRESS CHEMISTRY ANALYZER | PICCOLO ANALYZER | JJG | ABAXIS, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |