FDA Adverse Event Malfunction Summary report: N

T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY

MDR report key: 10264769 · Received July 13, 2020

Report

Report Number
3013756811-2020-71728
Event Type
Malfunction
Date Received
July 13, 2020
Date of Event
June 21, 2020
Report Date
July 13, 2020
Manufacturer
TANDEM DIABETES CARE
Product Code
OZO
UDI-DI
00853052007264
PMA / PMN Number
P180008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PER TANDEM'S USER GUIDE: DO NOT REUSE CARTRIDGES OR USE CARTRIDGES OTHER THAN THOSE MANUFACTURED BY TANDEM DIABETES CARE, INC. USE OF CARTRIDGES NOT MANUFACTURED BY TANDEM DIABETES CARE, INC. OR REUSE OF CARTRIDGES MAY RESULT IN OVER DELIVERY OR UNDER DELIVERY OF INSULIN. NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER REFILLED HIS CARTRIDGE AND SUBSEQUENTLY A MALFUNCTION ALARM OCCURRED. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 160 MG/DL. DURING TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT, THE MALFUNCTION ALARM WAS CLEARED, AND INSULIN DELIVERY WAS ABLE TO BE RESUMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
731249 T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND OZO TANDEM DIABETES CARE 1000096 00853052007264

Patients

Seq Age Sex Outcome Treatment
1 50 YR