THUNDERBEAT 5 MM, 35 CM, FRONT-ACTUATED GRIP TYPE S
Report
- Report Number
- 8010047-2020-04147
- Event Type
- Malfunction
- Date Received
- July 10, 2020
- Date of Event
- June 22, 2020
- Report Date
- October 29, 2020
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORP.
- Product Code
- GEI
- UDI-DI
- 04953170383540
- PMA / PMN Number
- K172610
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PHYSICIAN
Narratives
THE DEVICE HAS BEEN RETURNED AND AN EVALUATION COMPLETED FOR IT. ALSO, THOUGH THE DEVICE BROKE NO FRAGMENT FELL INTO THE PATIENT. THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THIS INFORMATION. KR910812 IS THE DEVICE PACKAGE LOT # AND THE ACTUAL LOT # ON THE DEVICE IS KR910678. THE CUSTOMER RETURNED THE DEVICE FOR EVALUATION FOR THE ISSUE OF THE PROBE BROKEN OFF. THE USER¿S COMPLAINT WAS CONFIRMED. THE DEVICE WAS ATTACHED TO THE USG-400/ESG-400 AND A PROBE CHECK WAS PERFORMED; THE DEVICE FAILED THE PROBE CHECK, ERROR CODE U509. BOTH SWITCHES WERE CHECKED AND FOUND BOTH SWITCHES ARE FUNCTIONAL. A VISUAL INSPECTION ON THE RECEIVED CONDITION WAS PERFORMED ON THE DEVICE; THERE IS SOME TISSUE BUILD UP. THE PTFE PAD (TEFLON PAD) WAS INSPECTED AND FOUND IT IS WORN WITH NO METAL EXPOSED. THE DISTAL END OF THE DEVICE WAS INSPECTED UNDER A MICROSCOPE AND FOUND THE PROBE DETACHED, MISSING PORTION NOT RETURNED. THE WIPER MOVEMENT IS NORMAL. THE CONSISTENCY OF MOVEMENT OF THE HANDLE AND JAW IS NORMAL. THE HANDLE LOAD IS NORMAL. THE ROTATION OF THE KNOB TORQUE IS NORMAL AND SMOOTH. THE DEVICE HISTORY RECORD REVIEW INDICATED NO ANOMALY IN MANUFACTURING THAT COULD BE ASSOCIATED WITH THE ISSUE. THE ROOT CAUSE FOR THE ISSUE CANNOT BE EXCLUSIVELY IDENTIFIED. HOWEVER BASED ON PAST INVESTIGATION RESULTS, THE PROBE BROKEN POSSIBLY OCCURRED DUE TO CONTACT WITH A SURGICAL INSTRUMENT OR THE NON-INSULATED AREA OF GRASPING SECTION. THE DETAILED STEP-BY-STEP SCENARIO CAN BE AS BELOW: <CONTACT WITH A SURGICAL INSTRUMENT> 1. DURING OUTPUT IN SEAL & CUT MODE, THE PROBE CAME IN CONTACT WITH HARD TISSUE, METAL OR A SURGICAL INSTRUMENT. CONSEQUENTLY, A SCRATCH WAS MADE ON THE PROBE AND THE ERROR OCCURRED. 2. THE PROBE RECEIVED AN OUTPUT LOAD IN SEAL & CUT MODE OR RECEIVED A LOAD WHEN GRASPING TISSUE. AS A RESULT, THE PROBE CRACKED FROM THE SCRATCH. 3. THE PROBE BROKE WHEN ADDED LOAD. <CONTACT WITH NON-INSULATED AREA OF GRASPING SECTION> 1. THE DISTAL END OF THE TISSUE PAD WAS WORN AWAY BECAUSE ¿SEAL & CUT¿ OUTPUT WAS ACTIVATED WHILE GRASPING NOTHING (INCLUDING THE CASE THE USER KEPT ACTIVATING AFTER CUTTING TISSUE). 2. THE TISSUE PAD WAS WORN AWAY, CAUSING THE NON-INSULATED OF THE GRASPING SECTION TO TOUCH THE PROBE. 3. ¿SEAL & CUT¿ OUTPUT WAS ACTIVATED UNDER THIS SITUATION, THEN THE SCRATCHES INDICATING THAT THE PROBE AND GRASPING SECTION WERE IN CONTACT WITH EACH OTHER WERE MADE. 4. THE PROBE RECEIVED AN OUTPUT LOAD IN SEAL & CUT MODE OR RECEIVED A LOAD WHEN GRASPING TISSUE. AS A RESULT, THE PROBE CRACKED FROM THE SCRATCH. 5. THE PROBE BROKE WHEN ADDED LOAD. THE INSTRUCTIONS FOR USE INCLUDES THE FOLLOWING STATEMENTS: ·DO NOT ACTIVATE OUTPUT IN SEAL & CUT MODE WHILE THE GRASPING SECTION IS CLOSED WITHOUT CONTACTING TISSUE OR VESSEL, OR ENSURING THAT TISSUE IS TRANSECTED. OTHERWISE, A LOCAL INCREASE OF THE TEMPERATURE DUE TO A FRICTION BETWEEN THE PROBE TIP AND THE GRASPING SECTION MAY RESULT IN VARIOUS FORMS OF DAMAGE IN THE PROBE TIP AND/OR THE TISSUE PAD, SUCH AS PREMATURE WEAR, BREAKAGE, DEFORMATION, AND/OR FALLING OFF INSIDE THE BODY CAVITY AND/OR PARTIAL SEPARATING. ¿ WHEN CUTTING AND VESSEL SEALING IS PERFORMED IN SEAL & CUT MODE, APPLY LIGHT TENSION ON THE TISSUE SO THAT USERS CAN CONFIRM IT IS TRANSECTED. ALSO, STOP ACTIVATION IMMEDIATELY AFTER TISSUE IS TRANSECTED. OTHERWISE, THE GRASPING SECTION, THE TISSUE PAD, OR THE PROBE TIP MAY BREAK AND FALL OFF, AND PARTIAL SEPARATING OF THE TISSUE PAD MAY OCCUR DUE TO A LOCAL INCREASE OF TEMPERATURE CAUSED BY THE FRICTION BETWEEN TISSUE PAD AND THE PROBE TIP DURING ACTIVATION. ¿ THE THUNDERBEAT INSTRUMENT SHOULD BE USED FOR SOFT TISSUE. DO NOT ACTIVATE OUTPUT WHILE GRASPING HARD TISSUE SUCH AS BONE OR HIGHLY CALCIFIED TISSUE, OR HARD OBJECTS SUCH AS METAL CLIPS, STAPLER, OR OTHER INSTRUMENTS (E.G., UTERINE MANIPULATOR, FORCEPS, AND OTHERS). OTHERWISE, IT MAY CAUSE THE PROBE TIP TO BE SCRATCHED OR COME INTO DIRECT CONTACT WITH THE METAL AREA OF THE GRASPING SECTION AS THE HEAT GENERATED BY THE FRICTION BETWEEN THE HARD OBJECT AND THE PROBE TIP COULD CAUSE WEAR/DEFORMING/SPLITTING/PROTRUDING/PARTIAL SEPARATING OF THE TISSUE PAD. IN TURN, THE PROBE MAY BREAK BEFORE DISPLAYING AN ERROR WINDOW OR GENERATING AN ALARM TONE. ¿ DO NOT GRASP OR LET THE PROBE TIP CONTACT HARD OBJECTS SUCH AS METAL CLIPS, STAPLER, OR OTHER INSTRUMENTS (E.G., UTERINE MANIPULATOR). ALSO, BE CAREFUL TO AVOID CONTACTING THE PROBE TIP WITH THOSE ACCIDENTALLY. PARTICULARLY DURING ACTIVATION, A SCRATCH ON THE PROBE TIP COULD OCCUR DUE TO ULTRASONIC VIBRATION, WHICH LEADS THE PROBE TIP TO BREAK AND FALL OFF INTO THE BODY CAVITY. IN ADDITION, THE HIGH-FREQUENCY (RF BIPOLAR) CURRENT FLOWS THROUGH THE METAL AND GENERATES SPARK DISCHARGE, WHICH MAY CAUSE BURNS AND DECREASE FUNCTIONALITIES.
DEVICE HAS NOT BEEN RETURNED FOR EVALUATION. THE CUSTOMER¿S COMPLAINT WAS NOT CONFIRMED. NO FINDINGS AVAILABLE. THE CAUSE COULD NOT BE DETERMINED. NO FURTHER INFORMATION WAS REPORTED. PER THE INSTRUCTIONS FOR USE, ¿THE THUNDERBEAT INSTRUMENT SHOULD BE USED FOR SOFT TISSUE. DO NOT ACTIVATE OUTPUT WHILE GRASPING HARD TISSUE SUCH AS BONE OR HIGHLY CALCIFIED TISSUE, OR HARD OBJECTS SUCH AS METAL CLIPS, STAPLER, OR OTHER INSTRUMENTS (E.G., UTERINE MANIPULATOR, FORCEPS, AND OTHERS). OTHERWISE, IT MAY CAUSE THE PROBE TIP TO BE SCRATCHED OR COME INTO DIRECT CONTACT WITH THE METAL AREA OF THE GRASPING SECTION AS THE HEAT GENERATED BY THE FRICTION BETWEEN THE HARD OBJECT AND THE PROBE TIP COULD CAUSE WEAR/DEFORMING/SPLITTING/PROTRUDING/PARTIAL SEPARATING OF THE TISSUE PAD.¿
THE CUSTOMER REPORTED TO OLYMPUS THAT DURING A THERAPEUTIC, LAP LIVER PROCEDURE, THE THUNDERBEAT BROKE. THERE WAS VISUAL DAMAGE TO THE PROBE UNIT. THE GENERATOR SETTINGS WERE 2/2. THERE WERE NO ERROR CODES DISPLAYED. THERE WAS NO CABLE DAMAGE OBSERVED. THE TRANSDUCER CORDS OR THE CORDS OF ANY OTHER MEDICAL DEVICE WERE NOT BUNDLED DURING USE. THE PHYSICIAN IS EXPERIENCED IN USING THE DEVICE. THE INTENDED PROCEDURE WAS COMPLETED WITH SIMILAR DEVICE. THERE WAS NO PATIENT INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 725305 | THUNDERBEAT 5 MM, 35 CM, FRONT-ACTUATED GRIP TYPE S | ULTRASONIC SURGICAL DEVICE | GEI | OLYMPUS MEDICAL SYSTEMS CORP. | TB-0535FCS | KR910678 | 04953170383540 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |