FDA Adverse Event Malfunction Summary report: N

SYRINGE 1.0ML 30GA 1/2IN UF 10BAG 500CS

MDR report key: 10255806 · Received July 9, 2020

Report

Report Number
2243072-2020-01045
Event Type
Malfunction
Date Received
July 9, 2020
Date of Event
July 1, 2020
Report Date
July 7, 2020
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
K170386
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. MEDICAL DEVICE EXPIRATION DATE: NA. INVESTIGATION SUMMARY: LEVEL A INVESTIGATION - COMPLAINT EVALUATION / COMPLAINT HISTORY CHECK FOR THE EVENT(S) THAT OCCURRED. SEVERITY: S_1__; OCCURRENCE: A COMPLAINT HISTORY CHECK WAS PERFORMED AND THIS IS THE 1ST RELATED COMPLAINT FOR LABEL INFORMATION MISSING (EXPIRATION DATE) ON LOT # 7093894. A REVIEW OF RISK MANAGEMENT 150RMN-0001-16 REVISION 13 INDICATES THAT THE POTENTIAL RISK OF THIS SPECIFIC REPORTED INCIDENT (SYRINGE, LABEL INFORMATION MISSING) WAS CAPTURED AND ADDRESSED APPROPRIATELY. INVESTIGATION SUMMARY: NO SAMPLES WERE RETURNED THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED AND THE ROOT CAUSE IS UNDETERMINED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. IF SAMPLES ARE RECEIVED IN THE FUTURE THE COMPLAINT WILL BE REOPENED FOR FURTHER INVESTIGATION. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 7093894. ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE ZERO (0) NOTIFICATIONS NOTED THAT PERTAINED TO THE COMPLAINT. THIS WAS MANUFACTURED BEFORE EXPIRATION DATES WERE ON PACKAGING. INVESTIGATION CONCLUSION: AS NO SAMPLES AND/OR PHOTO(S) WERE RECEIVED THE INVESTIGATION CONCLUDED: UNCONFIRMED: BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE AS NO SAMPLES OR PHOTOS WERE RETURNED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. ROOT CAUSE DESCRIPTION: ROOT CAUSE CANNOT BE DETERMINED AT THIS TIME AS THE ISSUE IS UNCONFIRMED AS NO SAMPLES OR PHOTOS WERE RETURNED. RATIONALE: BASED ON THE INVESTIGATION, NO ADDITIONAL INVESTIGATION AND NO CAPA IS REQUIRED AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF SYRINGES 1.0ML 30GA 1/2IN UF 10BAG 500CS EXPERIENCED MISSING LABEL INFORMATION WHICH WAS NOTED PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 328411 BATCH NO: 7093894. MEDICAL PROFESSIONAL REPORTED NO EXPIRATION DATE ON SEVERAL INSULIN SYRINGE BOXES. STATED EVERY BOX HAS THE SAME LOT #, CAT #, AND COPYRIGHT DATE OF 2013. DATE OF EVENT: UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
719657 SYRINGE 1.0ML 30GA 1/2IN UF 10BAG 500CS PISTON SYRINGE FMF BD MEDICAL - DIABETES CARE 7093894

Patients

Seq Age Sex Outcome Treatment
1 Other