BD 60ML SYRINGE LUER-LOK TIP
Report
- Report Number
- 1911916-2020-00636
- Event Type
- Malfunction
- Date Received
- July 9, 2020
- Date of Event
- June 17, 2020
- Report Date
- July 10, 2020
- Manufacturer
- BD MEDICAL (BD WEST) MEDICAL SURGICAL
- Product Code
- FMF
- UDI-DI
- 30382903096535
- PMA / PMN Number
- K980987
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION: ONE PHOTO WAS PROVIDED FOR EVALUATION. IT SHOWS A PACKAGING BOX LABEL WITH 60 ML AND LOT# 8165956 AND SHOWS ANOTHER BOX LABEL OF 50 ML LOT# 9240280. ADDITIONALLY, IT SHOWS TWO PACKAGING BLISTERS ONE 60 ML AND ANOTHER ONE 50 ML. FROM THE PHOTOS, WE CAN CONCLUDE THAT THE CUSTOMER HAS A BOX(ES) LABELED AS 60 ML WITH PRODUCT 60 ML FROM THE LOT 8165956; THE CUSTOMER ALSO HAS A BOX(ES) LABELED AS 50 ML WITH 50 ML SYRINGES LOT# 9240280. BASED ON THE PHOTOS PROVIDED, WE CONCLUDE THAT AT A CERTAIN POINT OF THE SUPPLY CHAIN BOXES WITH 50 ML AND 60 ML SYRINGES WERE PUT TOGETHER. THE MANUFACTURING SITE IS RULED OUT SINCE THESE LOTS WERE PRODUCED WITH OVER A YEAR APART FROM EACH OTHER. IT IS UNKNOWN WHERE IT COULD HAVE HAPPENED. BASED ON THE INVESTIGATION AND WITH THE PHOTO PROVIDED, THE SYMPTOM REPORTED BY THE CUSTOMER IS CONFIRMED. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED WITH ZERO DEFECTS FOUND. NO QUALITY NOTIFICATIONS WERE WRITTEN FOR THIS BATCH, NOR FOR THE ASSOCIATED ASSEMBLY BATCHES.
IT WAS REPORTED THAT LABELS ON PACKAGES SAY 60ML AND OTHERS HAVE 50ML WITH A BD 60ML SYRINGE LUER-LOK¿ TIP. THIS OCCURRED ON 15 SEPARATE OCCASIONS PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT SOME PACKAGE LABELS HAVE 60ML AND OTHERS HAVE 50ML.
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
IT WAS REPORTED THAT LABELS ON PACKAGES SAY 60ML AND OTHERS HAVE 50ML WITH A BD 60ML SYRINGE LUER-LOK¿ TIP. THIS OCCURRED ON 15 SEPARATE OCCASIONS PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT SOME PACKAGE LABELS HAVE 60ML AND OTHERS HAVE 50ML.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 719152 | BD 60ML SYRINGE LUER-LOK TIP | PISTON SYRINGE | FMF | BD MEDICAL (BD WEST) MEDICAL SURGICAL | 309653 | 9240280 | 30382903096535 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |