MEDCARE EXTRA LARGE CARE SLING W/HEAD SUPPORT - KNIT

MDR report key: 10251386 · Received July 9, 2020

Report

Report Number
3007802293-2020-00011
Event Type
Malfunction
Date Received
July 9, 2020
Date of Event
June 19, 2020
Report Date
October 16, 2020
Manufacturer
HANDICARE USA INC.
Product Code
FSA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

IMAGES OF THE SLING WERE REVIEWED AS THE SLING WAS NOT RETURNED. FOR ALL FOUR STRAPS TO FAIL SIMULTANEOUSLY THERE WOULD HAVE TO BE SOME SORT OF EXTERNAL PRESSURE OR CUTTING DUE TO THE HOOKS OF THE CARRY BAR BEING ROUGH OR PITTED. HOWEVER, PICTURES OF THE FLOOR LIFT WITH CARRY BAR WERE NOT PROVIDED TO CONFIRM THIS THEORY. THIS SLING FAMILY HAS BEEN LOAD TESTED AT 1500 LBS 1.5X THE SAFE WORKING LOAD OF 1000LBS. THE SLING WAS REMOVED FROM SERVICE.

Additional Manufacturer Narrative · 1

WE ARE AWAITING FURTHER DETAILS AND THE PRODUCT FROM THE FACILITY. A FOLLOW-UP REPORT WILL BE FILED WITH THE RESULTS OF THE INVESTIGATION.

Description of Event or Problem · 1

A USER WAS BEING TRANSFERRED BY TWO NURSES AND TWO CNAS. FOUR STRAPS OF THE SLING BROKE. THE USER FELL TO THE FLOOR AND IS IN PAIN. NO INJURIES ARE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
713115 MEDCARE EXTRA LARGE CARE SLING W/HEAD SUPPORT - KNIT LIFTING SLING FSA HANDICARE USA INC. 400034