FDA Adverse Event Malfunction Summary report: N

BD ECLIPSE NEEDLE

MDR report key: 10243051 · Received July 7, 2020

Report

Report Number
8041187-2020-00388
Event Type
Malfunction
Date Received
July 7, 2020
Date of Event
June 8, 2020
Report Date
July 30, 2020
Manufacturer
BECTON DICKINSON MEDICAL (SINGAPORE)
Product Code
FMI
UDI-DI
30382903057659
PMA / PMN Number
K161170
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: ONE PHOTO SAMPLE WAS RECEIVED BY OUR QUALITY TEAM FOR EVALUATION. FROM THE PHOTO, SIGNS OF THE SAFETY SHIELD BEING BROKE OFF AND MISSING FROM THE UNIT PACKAGE WAS OBSERVED. THE CUSTOMER EXPERIENCE WAS CONFIRMED FROM THE PHOTO EVALUATION. A REVIEW OF THE INTERNAL MANUFACTURING DEVICE RECORDS AND RAW MATERIAL HISTORY FILES FOR THE REPORTED LOT NUMBERS WAS PERFORMED AND NO RECORDED QUALITY PROBLEMS OR REJECTIONS TO THIS INCIDENT WERE FOUND. THE PRIMARY PACKAGING PROCESS WAS REVIEWED. THERE IS A SAFETY SHIELD PRESENCE SENSOR TO DETECT FOR MISSING SAFETY SHIELD AND AUTO-REJECT THE PART. THE SENSOR WAS CHALLENGED AND ABLE TO AUTO-REJECT THE PART. THE SAFETY SHIELD COULD HAVE EXPERIENCED SIGNS OF BREAKAGE DURING EXERCISING MOTION AND DROPPED OFF WHEN MOVING TO THE NEXT STATION. REVIEWED THE MOLDING FIRST PIECE REPORT AND THE ROUNDNESS AND ROD OD MEASUREMENTS WERE WITHIN SPECIFICATION. ROOT CAUSE COULD NOT BE DETERMINED FOR SAFETY SHIELD BROKE OFF AS NO SAMPLE IS RETURNED FOR INVESTIGATION. THERE IS A VISION SYSTEM AT THE PRIMARY PACKAGING MACHINE TO CHECK FOR MISSING SAFETY SHIELD. UPON DETECTED MISSING SHIELD, THE VISION SYSTEM WILL TRIGGER THE MACHINE TO STOP. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD ECLIPSE¿ NEEDLE SAFETY SHIELD WAS FOUND DETACHED BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WE CONTINUE TO FIND UNACCEPTABLE UNITS (MISSING SAFETY SHIELD) IN LOT NUMBER 8171369."

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BD ECLIPSE¿ NEEDLE SAFETY SHIELD WAS FOUND DETACHED BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WE CONTINUE TO FIND UNACCEPTABLE UNITS (MISSING SAFETY SHIELD) IN LOT NUMBER 8171369."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
707104 BD ECLIPSE NEEDLE HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON DICKINSON MEDICAL (SINGAPORE) 305765 8171369 30382903057659

Patients

Seq Age Sex Outcome Treatment
1 Other