FDA Adverse Event Malfunction Summary report: N

CADD

MDR report key: 10242987 · Received July 7, 2020

Report

Report Number
3012307300-2020-06695
Event Type
Malfunction
Date Received
July 7, 2020
Report Date
August 17, 2020
Manufacturer
ST PAUL
Product Code
FRN
PMA / PMN Number
K040899
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

RETURNED DEVICE WAS RECEIVED IN GOOD PHYSICAL CONDITION. NO EVIDENCE OF REPORTED PROBLEM IN EVENT LOG WAS FOUND. DURING THE EVALUATION OF THE DEVICE, NO FAULT WAS FOUND WITH THE PUMP. THREE DELIVERY TESTS WERE PERFORMED AND WERE FOUND TO BE WITHIN SPECIFICATIONS. THE DEVICE HISTORY RECORD WAS REVIEWED AND SHOWED THAT THIS DEVICE MET ALL MANUFACTURING SPECIFICATION FOR PRODUCT RELEASED FOR DISTRIBUTION. NO ISSUES WERE IDENTIFIED THAT WOULD HAVE IMPACTED THIS EVENT.

Description of Event or Problem · 0

ORACLE RO 1072008: FAIL ACCURACY - 12.45%. ADDITIONAL INFORMATION RECEIVED ON 26-JUN-2020: NO PATIENT INVOLVEMENT. EVENT DETAILS UPDATED.

Additional Manufacturer Narrative · 1

RETURNED DEVICE WAS RECEIVED IN DECENT PHYSICAL CONDITION. THE DEVICE HAD A CRACKED BOTTOM ASE IN THE BATTERY BAY AND A RIGHT PLUNGER CASE. NO EVIDENCE OF REPORTED PROBLEM IN EVENT LOG WAS FOUND. DURING THE EVALUATION OF THE DEVICE, THE SPEAKER WAS MAKING A RATTLING NOISE DURING THE POWER UP PROCESS AND TESTING THE SPEAKER. THERE WAS DEBRIS FOUND STUCK INSIDE THE SPEAKER. THIS WAS LIKELY CAUSED BY NORMAL WEAR AS THE PUMP IS OVER 14 YEARS OLD. THE DEVICE HISTORY RECORD WAS REVIEWED AND SHOWED THAT THIS DEVICE MET ALL MANUFACTURING SPECIFICATION FOR PRODUCT RELEASED FOR DISTRIBUTION. NO ISSUES WERE IDENTIFIED THAT WOULD HAVE IMPACTED THIS EVENT.

Description of Event or Problem · 1

INFORMATION WAS RECEIVED INDICATING THAT A SMITHS MEDICAL MEDFUSION SYRINGE PUMP PRESENTED WITH AN AUDIBLE ALARM. THERE WERE NO REPORTED ADVERSE EVENTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
705183 CADD PUMP, INFUSION FRN ST PAUL 21-6400-51

Patients

Seq Age Sex Outcome Treatment
1