INRATIO
Report
- Report Number
- 2954730-2008-00146
- Event Type
- Malfunction
- Date Received
- March 27, 2008
- Date of Event
- March 3, 2008
- Report Date
- March 25, 2008
- Manufacturer
- HEMOSENSE SJ
- Product Code
- JPA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
INRATIO PRECISION DATA PROVIDED BY END-USER LOT 070299: PT#1: FIRST TEST INR =5.3(METER 1), SECOND TEST INR = 4.3(METER 2), MEAN = 4.8; SD = 0.7; %CV = 15%. THE %CV IS LESS THAN OR EQUAL TO 20%. PER INTERNAL PROCEDURE, THE PRECISION PASSES THE CRITERIA FOR PRECISION. THE TEST IS CONSIDERED PRECISE AND NO FURTHER TESTING IS REQUIRED AT THIS TIME. INRATIO PRECISION DATA PROVIDED BY END-USER LOT 070299: PT#2: FIRST TEST INR = 2.2(METER 1), SECOND TEST INR = 1.1(METER 2), MEAN = 1.65; SD = 0.77; %CV = 47%. THE %CV IS GREATER THAN 20%. PER INTERNAL PROCEDURE, THE PRECISION FAILED THE CRITERIA FOR PRECISION. PRODUCTS WILL BE TESTED WHEN RETURNED.
CALLER ALLEGED IMPRECISION WITH INRATIO METER. RESULTS AS FOLLOWS: PT#1: FIRST TEST INR = 5.3(METER 1), SECOND TEST INR = 4.3(METER 2); PT#2: FIRST TEST INR = 2.2(METER 1), SECOND TEST INR = 1.1(METER 2).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INRATIO | PROTHROMBIN TIME TEST | JPA | HEMOSENSE SJ | 070299 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |