LIFEPAK 20 DEFIBRILLATOR/MONITOR
MDR report key: 1024181
·
Received March 26, 2008
Report
- Report Number
- 3015876-2008-00222
- Event Type
- Malfunction
- Date Received
- March 26, 2008
- Date of Event
- February 4, 2008
- Report Date
- February 25, 2008
- Manufacturer
- PHYSIO-CONTROL, INC.
- Product Code
- LDD
- PMA / PMN Number
- K012274
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
PHYSIO-CONTROL EVALUATED THE DEVICE AND OBSERVED THAT IT WOULD INTERMITTENTLY POWER OFF BY ITSELF. PHYSIO REPLACED THE POWER SUPPLY ASSEMBLY AND THEN OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE. PHYSIO-CONTROL FURTHER EVALUATED THE REPLACED ASSEMBLY AND DETERMINED THAT THE ROOT CAUSE FOR THE REPORTER FAILURE WAS A LEAKY FILTER, FL4.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE WOULD POWER OFF BY ITSELF. THERE WAS NO PT INVOLVEMENT REPORTED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEPAK 20 DEFIBRILLATOR/MONITOR | LDD | PHYSIO-CONTROL, INC. | 20 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |