LIFEPAK 20 DEFIBRILLATOR/MONITOR

MDR report key: 1024181 · Received March 26, 2008

Report

Report Number
3015876-2008-00222
Event Type
Malfunction
Date Received
March 26, 2008
Date of Event
February 4, 2008
Report Date
February 25, 2008
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
LDD
PMA / PMN Number
K012274
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL EVALUATED THE DEVICE AND OBSERVED THAT IT WOULD INTERMITTENTLY POWER OFF BY ITSELF. PHYSIO REPLACED THE POWER SUPPLY ASSEMBLY AND THEN OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE. PHYSIO-CONTROL FURTHER EVALUATED THE REPLACED ASSEMBLY AND DETERMINED THAT THE ROOT CAUSE FOR THE REPORTER FAILURE WAS A LEAKY FILTER, FL4.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WOULD POWER OFF BY ITSELF. THERE WAS NO PT INVOLVEMENT REPORTED WITH THIS EVENT.