FDA Adverse Event
Malfunction
Summary report: N
COLLEAGUE CX VOLUMETRIC INFUSION PUMP.
MDR report key: 1024002
·
Received April 4, 2008
Report
- Report Number
- 6000001-2007-02288
- Event Type
- Malfunction
- Date Received
- April 4, 2008
- Date of Event
- January 16, 2007
- Report Date
- January 16, 2007
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Removal / Correction Number
- 6000001-7/20/05-015-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
EVALUATION SUMMARY: EVALUATION WAS PERFORMED AND THE REPORTED CONDITION WAS CONFIRMED. INSPECTION OF THE PUMP REVEALED DEFECTIVE USER INTERFACE MODULE PRINTER CIRCUIT BOARD (UIM PCB) CAUSED FAILURE CODE 804:22. THE UIM PCB WAS REPLACED. THE PUMP WAS TESTED FOR PROPER OPERATION AND PUMP FOUND TO FUNCTION PROPERLY. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS CURRENTLY BEING INVESTIGATED.
Description of Event or Problem · 1
DURING SERVICE BY BAXTER, FAILURE CODE 804:22 WAS FOUND. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION OR CONTACT WAS AVAILABLE.