FDA Adverse Event Malfunction Summary report: N

COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP.

MDR report key: 1023998 · Received April 4, 2008

Report

Report Number
6000001-2007-02482
Event Type
Malfunction
Date Received
April 4, 2008
Date of Event
January 12, 2007
Report Date
January 12, 2007
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Removal / Correction Number
CONTINUED ON H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: EVALUATION WAS COMPLETED AND THE REPORTED CONDITION OF DEPLETED MAIN BATTERIES WAS CONFIRMED. INSPECTION OF THE DEVICE REVEALED POTENTIALLY DAMAGED BATTERIES. THE MAIN BATTERIES AND THE BATTERY HARNESS WERE REPLACED AND THE BAXTER TECHNICIAN TESTED THE DEVICE. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS CURRENTLY BEING INVESTIGATED UNDER CAPA.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED FOR SERVICE. DURING SERVICE BY BAXTER, DEPLETED MAIN BATTERIES WERE FOUND. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION OR CONTACT WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP. INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. NA NA