FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX VOLUMETRIC INFUSION PUMP.

MDR report key: 1023964 · Received April 4, 2008

Report

Report Number
6000001-2007-03739
Event Type
Malfunction
Date Received
April 4, 2008
Date of Event
December 1, 2006
Report Date
December 26, 2006
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Removal / Correction Number
CONTINUED ON H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE REPORTED CONDITION OF THE PUMP NEEDING NEW BATTERIES FOR UNKNOWN REASONS WAS CONFIRMED. INSPECTION OF THE DEVICE FOUND THAT THE PUMP'S BATTERIES WERE DAMAGED WITH 26 BATTERY DISCHARGES BELOW THE ALARM THRESHOLD, THEREFORE, THEY WERE REPLACED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS CURRENTLY BEING INVESTIGATED UNDER CAPA.

Description of Event or Problem · 1

THE FACILITY REPORTED A PUMP THAT NEEDS A NEW BATTERY FOR UNKNOWN REASONS. IT IS UNKNOWN WHEN THIS OCCURRED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.