FDA Adverse Event Malfunction Summary report: N

FLO-GARD 6301 VOLUMETRIC INFUSION PUMP

MDR report key: 1023937 · Received April 4, 2008

Report

Report Number
6000001-2007-04611
Event Type
Malfunction
Date Received
April 4, 2008
Date of Event
December 1, 2006
Report Date
December 26, 2006
Manufacturer
BAXTER HEALTHCARE (SG)
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE FACILITY REPRESENTATIVE REPORTED AN OCCLUSION ALARM. INFORMATION WAS NOT PROVIDED REGARDING WHETHER OR NOT THE PROBLEM OCCURED DURING A PATIENT INFUSION. EVALUATION SUMMARY: THE PUMP WAS EVALUATED BY A BAXTER SERVICE TECHNICIAN. INSPECTION OF THE DEVICE FOUND THAT THE OCCLUSION ALARM WAS CAUSED BY BROKEN DOOR HINGES ON CHANNEL 2. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS CURRENTLY BEING INVESTIGATED UNDER CAPA.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED FOR SERVICE. DURING SERVICE BY BAXTER, BROKEN DOOR HINGES WERE FOUND. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLO-GARD 6301 VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE (SG) NA NA

Patients

Seq Age Sex Outcome Treatment
1