COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP.
Report
- Report Number
- 6000001-2007-01855
- Event Type
- Malfunction
- Date Received
- April 4, 2008
- Date of Event
- December 1, 2006
- Report Date
- December 26, 2006
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Removal / Correction Number
- CONTINUED ON H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
EVALUATION SUMMARY: THE REPORTED CONDITION OF FAILURE CODE 570 WAS CONFIRMED BY THE BAXTER REPAIR TECHNICIAN, DURING ON SITE SERVICE TESTING. INSPECTION OF THE DEVICE REVEALED POTENTIALLY DEPLETED MAIN BATTERIES, WHICH WERE REPLACED. THE DEVICE WAS TESTED BY THE REPAIR TECHNICIAN.REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS CURRENTLY BEING INVESTIGATED UNDER CAPA.
THE FACILITY REPORTED AN INFUSION PUMP WITH A FAILURE CODE 570. INFORMATION WAS NOT PROVIDED REGARDING WHETHER OR NOT THE FAILURE OCCURED DURING AN INFUSION. THE HOSPITAL REPRESENTATIVE STATED THAT THERE HAVE BEEN NO REPORTS OF ANY PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS KNOWN.