FDA Adverse Event Malfunction Summary report: N

COLLEAGUE VOLUMETRIC INFUSION PUMP

MDR report key: 1023920 · Received April 4, 2008

Report

Report Number
6000001-2007-01905
Event Type
Malfunction
Date Received
April 4, 2008
Date of Event
December 1, 2006
Report Date
December 26, 2006
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NM, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE CONDITION OF FAIL CODE 812:02 WAS CONFIRMED IN THE EVENT HISTORY. CORROSION ON PUMP HEAD MODULE PRINTED CIRCUIT BOARD AT R131 AND C66 CAUSED THIS FAIL CODE. THE PUMP HEAD MODULE ASSEMBLY WAS REPLACED.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED FAILURE CODE 812:02. INFORMATION WAS NOT PROVIDED REGARDING WHETHER OR NOT THE FAILURE OCCURED DURING A PATIENT INFUSION. THE HOSPITAL REPRESENTATIVE STATED THAT THERE WER NO REPORTS OF PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD.

Patients

Seq Age Sex Outcome Treatment
1