FDA Adverse Event Malfunction Summary report: N

COLLEAGUE 3 VOLUMETRIC INFUSION PUMP

MDR report key: 1023893 · Received April 4, 2008

Report

Report Number
6000001-2007-00890
Event Type
Malfunction
Date Received
April 4, 2008
Date of Event
December 1, 2006
Report Date
December 11, 2006
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Removal / Correction Number
6000001-7/20/05-015-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 09 2007. EVALUATION SUMMARY:THE CONDITION OF FAIL CODE 703 WAS CONFIRMED. THIS FAIL CODE WAS ASSOCIATED WITH THE USER INTERFACE MODULE COMMUNICATION WITH THE PUMP HEAD MODULE. THE USER INTERFACE MODULE WAS DEFECTIVE AND WAS REPLACED. THIS ISSUE IS CURRENTLY BEING INVESTIGATED UNDERB CAPA. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED AN INFUSION PUMP WITH FAILURE CODE 703 DURING BIOMED TESTING. THE HOSPITAL REPRESENTATIVE STATED THAT THERE WERE NO REPORTS OF PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1