FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX VOLUMETRIC INFUSION PUMP

MDR report key: 1023886 · Received April 4, 2008

Report

Report Number
6000001-2007-00675
Event Type
Malfunction
Date Received
April 4, 2008
Date of Event
December 1, 2006
Report Date
December 11, 2006
Manufacturer
BAXTER HEALTHCARE PTE LTD
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 08 2007. EVALUATION SUMMARY:THE INFUSION PUMP WAS TESTED FOR FLOW ACCURACY AT 100 ML/HOUR, THE INFUSION PUMP FAILED THE ACCURACY TEST. THE SPECIFICATION OF THIS DEVICE IS +/-5%. THE REPORTED CONDITION OF UNDER-INFUSION WAS CONFIRMED AND IT WAS DUE TO THE PUMP FAILING THE PRE-ACCURACY TEST, THE PUMP HEAD MODULE WAS REPLACED.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED,THAT THE DEVICE WAS UNDER-INFUSING. THE EVENT OCCURRED DURING BIO-MED TESTING. THE HOSPITAL REPRESENTATIVE STATED,THAT THERE WERE NO REPORTS OF PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE. THE REPORTED CONDITION WAS CONFIRMED DURING SERVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CX VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE LTD NA NA

Patients

Seq Age Sex Outcome Treatment
1