FDA Adverse Event
Malfunction
Summary report: N
COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP
MDR report key: 1023882
·
Received April 4, 2008
Report
- Report Number
- 6000001-2007-02305
- Event Type
- Malfunction
- Date Received
- April 4, 2008
- Date of Event
- December 29, 2006
- Report Date
- December 29, 2006
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HI, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
EVALUATION SUMMARY: THE INFUSION PUMP WAS TESTED FOR FLOW ACCURACY AT 100 ML/HOUR, THE INFUSION PUMP FAILED THE ACCURACY TEST. THE SPECIFICATION OF THIS DEVICE IS +/-5%. THE CONDITION OF THE CHANNEL C PUMP HEAD MODULE FAILING THE 1 HOUR ACCURACY TEST WAS CONFIRMED. THE CHANNEL C PUMP HEAD MODULE WAS REPLACED.
Description of Event or Problem · 1
THE FACILITY RETURNED THE DEVICE FOR SERVICE. DURING SERVICE THE BAXTER REPAIR TECHNICIAN NOTED THE CHANNEL C PUMP HEAD MODULED FAILED THE 1 HOUR ACCURACY TEST. THE HOSPITAL REPRESENTATIVE STATED THAT THERE HAVE BEEN NO REPORTS OF ANY PATIENT INCIDENT INVOLVING THIS PUMP. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE. LTD. | NA | NA |