FDA Adverse Event Malfunction Summary report: N

COLLEAGUE 3 VOLUMETRIC INFUSION PUMP

MDR report key: 1023876 · Received April 4, 2008

Report

Report Number
6000001-2007-01136
Event Type
Malfunction
Date Received
April 4, 2008
Date of Event
December 1, 2006
Report Date
December 11, 2006
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Removal / Correction Number
6000001-3/15/05-007-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE REPORTED CONDITION OF FAILURE CODE 810:04 WAS CONFIRMED IN THE PUMP'S EVENT HISTORY FILE DURING PRODUCT EVALUATION. INSPECTION OF THE DEVICE FOUND THAT THIS FAILURE CODE WAS CAUSED BY AN OUT OF CALIBRATION AIR IN LINE PRINTED CIRCUIT BOARD ON CHANNEL B. THE AIR IN LINE PRINTED CIRCUIT BOARD ON CHANNEL B WAS RECALIBRATED.

Description of Event or Problem · 1

THE FACILITY REPORTED A PUMP WITH FAILURE CODE 810:04. THIS PROBLEM WAS IDENTIFIED DURING BIO-MED TESTING. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.