FDA Adverse Event
Malfunction
Summary report: N
PRECISION XCEED
MDR report key: 1023853
·
Received April 2, 2008
Report
- Report Number
- 2954323-2008-01295
- Event Type
- Malfunction
- Date Received
- April 2, 2008
- Date of Event
- March 5, 2008
- Report Date
- April 2, 2008
- Manufacturer
- ABBOTT DIABETES CARE LIMITED UK
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE CUSTOMER'S PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE INVESTIGATION RESULTS ARE AVAILABLE.
Description of Event or Problem · 1
A HOSPITAL REPORTED OBTAINING TWO SETS OF HIGH READINGS ON A PRECISION XTRA METER. THE CUSTOMER REPORTED THEY OBTAINED A READING OF 500 MG/DL COMPARED TO A LAB READING OF 74 MG/DL AND A READING OF 390 MG/DL COMPARED TO A LAB READING OF 181 MG/DL, ALL READINGS WERE TAKEN WITHIN A 10 MINUTE TIMEFRAME. WHEN PLOTTED ON A PARKES ERROR GRID THE RESULTS FELL INTO THE "D" ZONE AND "C" ZONE RESPECTIVELY, SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF A SERIOUS INJURY, DEATH OR PERMANENT IMPAIRMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PRECISION XCEED | BLOOD GLUCOSE METER | NBW | ABBOTT DIABETES CARE LIMITED UK | NI | 42359 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI |