FDA Adverse Event Malfunction Summary report: N

SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA

MDR report key: 1023825 · Received April 2, 2008

Report

Report Number
2023826-2008-00458
Event Type
Malfunction
Date Received
April 2, 2008
Date of Event
March 12, 2008
Report Date
March 12, 2008
Manufacturer
STAAR SURGICAL
Product Code
HQL
PMA / PMN Number
P900048
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

EVAL: RESULTS: A VISUAL INSPECTION OF THE RETURNED PRODUCT SHOWS BOTH HAPTICS ARE TORN OFF INTO THE OPTIC OF THE LENS AND ARE MISSING. THE OPTIC HAS TEARS IN IT. THERE IS CLEAR SURGICAL RESIDUE ON THE LENS. IT SHOULD BE NOTED THAT THE INJECTOR AND CARTRIDGE WERE NOT RETURNED FOR EVAL. CONCLUSIONS - AN INVESTIGATION WAS OPENED TO EVALUATE A COMPLAINT TREND ASSOCIATED WITH LENS TEARS THAT WAS ORIGINALLY IDENTIFIED IN 2005. POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND POSSIBLE HANDLING ERRORS BY THE CUSTOMER. TO ADDRESS DELIVERY SYSTEM ISSUES, ALL STAGES IN THE MANUFACTURING OF THE INJECTORS AND CARTRIDGES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. TO ADDRESS HANDLING ERRORS, ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS.

Description of Event or Problem · 1

THE REPORTER STATED THAT THE SURGEON WAS INSERTING A THREE PIECE SILICONE LENS MODEL AQ2010V AND THE LENS TORE. THE LENS WAS REMOVED WITHOUT ENLARGING THE INCISION OR SUTURE REQUIRED. NO PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA INTRAOCULAR LENS HQL STAAR SURGICAL AQ2010V NA

Patients

Seq Age Sex Outcome Treatment
1 INJECTOR MODEL: MSI-TM| CARTRIDGE MODEL: AQ CARTRIDGE-FP