FDA Adverse Event Injury Summary report: N

EQUATE BLOOD GLUCOSE STRIPS

MDR report key: 10238045 · Received July 7, 2020

Report

Report Number
3005862821-2020-00037
Event Type
Injury
Date Received
July 7, 2020
Date of Event
May 26, 2020
Report Date
June 17, 2020
Manufacturer
OK BIOTECH CO., LTD.
Product Code
CGA
PMA / PMN Number
K162430
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

NO NON-CONFORMITY WAS FOUND AFTER OKB REVIEWED ALL MANUFACTURING AND QC RECORDS OF STRIPS (LOT#: U19100412). STRIPS WERE MANUFACTURED ON 10/04/2019 AND WILL EXPIRE IN 10/2021. BECAUSE THE SUSPECTED STRIPS WERE NOT RETURNED TO OKB, THE RETAINED STRIPS (LOT#: U19100412) FROM OKB'S WAREHOUSE WERE USED TO RE-TEST WITH ONETOUCH METER (CODE: 49) AND CONTROL SOLUTION (LEVEL LOW: BATCH# 8AH1A95, EXP. BY DEC., 2020; LEVEL HIGH: BATCH# 8AH3A15, EXP. BY DEC., 2020). TESTING RESULTS (LEVEL LOW: 46/43; LEVEL HIGH: 259/256) MET THE ACCEPTANCE CRITERIA (LEVEL LOW: 30~80; LEVEL HIGH: 230~340), AND DESICCANTS OF STRIP VIAL ARE FUNCTIONAL. THE ROOT CAUSE OF THE COMPLAINT IS UNABLE TO BE VERIFIED WITHOUT THE SUSPECTED STRIPS AND MORE USER'S INFORMATION. THEREFORE, THE MATTER HAS TO BE CLOSED OUT IF NO FURTHER ACTION OR INFORMATION FROM THE USER.

Description of Event or Problem · 1

CALLER STATED THAT MEDICAL ATTENTION WAS SOUGHT ON (B)(6) 2020 DUE TO HER EQUATE TEST STRIPS GIVING A LOW RESULT. THE CALLER STATED THAT SHE GAVE HER SON INSULIN BASED OFF THE RESULT SHE RECEIVED WHEN SHE TESTED HIS BLOOD GLUCOSE WITH THE EQUATE TEST STRIPS. SHE STATED THAT THIS CAUSED HER TO HAVE TO TAKE HER SON TO THE HOSPITAL. WE HAVE ATTEMPTED TO CONTACT THE CALLER TO OBTAIN MORE INFORMATION AND LEFT MULTIPLE MESSAGES AND THE CALLER HAS NOT CALLED BACK TO PROVIDE US WITH MORE INFORMATION. WE DO NOT HAVE ANY INFORMATION ON THE END-USER. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
706648 EQUATE BLOOD GLUCOSE STRIPS BLOOD GLUCOSE TEST STRIPS CGA OK BIOTECH CO., LTD. U19100412

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization