BD PLASTIPAK SYRINGE LUER SLIP
Report
- Report Number
- 3003916417-2020-00192
- Event Type
- Malfunction
- Date Received
- July 6, 2020
- Date of Event
- June 15, 2020
- Report Date
- August 3, 2020
- Manufacturer
- BECTON DICKINSON IND. CIRURGICAS LTDA
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H.6. INVESTIGATION SUMMARY A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 9326884 MAINTENANCE RECORD ANALYSIS AND QUALITY NOTIFICATION ANALYSIS AND NO DEVIATION WERE FOUND FOR THIS BATCH. PHOTOS WERE SENT BY THE CUSTOMER AND IT WAS POSSIBLE TO OBSERVE TOP WEB PACKAGE MIX, SYRINGE LL X LS. ACCORDING THE EVALUATION OF BATCH HISTORY AND PROCESS THE POTENTIAL CAUSES FOR THE OCCURRENCE ARE: 1 - OPERATIONAL FAILURE DURING LINE SETUP AND LINE CLEARANCE, LEADING TO THE TOP WEB MIX. 2 - FAILURE AT LOGISTIC STEP OF MATERIAL SEPARATION, LEADING TO THE TOP WEB MIX AS ACTIONS FOR THE INCIDENT: 1 - THE OPERATORS OF THE LINE INVOLVED, AND THE LOGISTICS TEAM WILL BE NOTIFIED OF THE OCCURRENCE. 2 - THE MAINTENANCE ORDER WAS OPENED TO REPLACE THE PAPER CONTROL SENSORS OF THE PACKAGING - ADDITIONALLY, THE INCIDENT REPORTED FROM THIS COMPLAINT WILL BE MONITORED FOR TREND EVALUATION. H3 OTHER TEXT : SEE H.10.
IT WAS REPORTED THAT THE BOX OF BD PLASTIPAK¿ SYRINGES WITH THE LUER SLIP CONTAINED LUER -LOK TIP SYRINGES INSTEAD, BUT THE BATCH LISTED ON THE LABEL FOR BOTH CATALOG NUMBERS WAS THE SAME. THIS WAS NOTICED PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM PORTUGUESE TO ENGLISH: "PRODUCTS IDENTIFIED IN THE LUER PLASTIPAK SYRINGE BOX - SLIP 20 ML, REF 990173 LOT: 9326884 VAL 11-2024, BUT INSIDE CONTAINS THE LUER -LOK TIP SYRINGE REF990687. QUANTITY 750 UNITS. ADDITIONAL INFORMATION: IN TELEPHONE CONTACT THE CUSTOMER DESCRIBES THE LABEL THAT IDENTIFIES THE BOX AND IT IS UNDERSTOOD THAT IT APPEARS TO BE THE LABEL ISSUED BY THE FACTORY. IN ADDITION, SHE NOTES THAT THE BATCH OF BOTH CATALOGS IS THE SAME (9326884)."
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. INITIAL REPORTER PHONE#: (B)(6). INITIAL REPORTER ADDRESS 1: (B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT THE BOX OF BD PLASTIPAK¿ SYRINGES WITH THE LUER SLIP CONTAINED LUER -LOK TIP SYRINGES INSTEAD, BUT THE BATCH LISTED ON THE LABEL FOR BOTH CATALOG NUMBERS WAS THE SAME. THIS WAS NOTICED PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM PORTUGUESE TO ENGLISH: "PRODUCTS IDENTIFIED IN THE LUER PLASTIPAK SYRINGE BOX - SLIP 20 ML, REF 990173 LOT: 9326884 VAL 11-2024, BUT INSIDE CONTAINS THE LUER -LOK TIP SYRINGE REF990687. QUANTITY 750 UNITS. ADDITIONAL INFORMATION: IN TELEPHONE CONTACT THE CUSTOMER DESCRIBES THE LABEL THAT IDENTIFIES THE BOX AND IT IS UNDERSTOOD THAT IT APPEARS TO BE THE LABEL ISSUED BY THE FACTORY. IN ADDITION, SHE NOTES THAT THE BATCH OF BOTH CATALOGS IS THE SAME (9326884).".
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 696455 | BD PLASTIPAK SYRINGE LUER SLIP | PISTON SYRINGE | FMF | BECTON DICKINSON IND. CIRURGICAS LTDA | 9326884 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |