FDA Adverse Event Malfunction Summary report: N

TAXUS EXPRESS2 PACLITAXEL-ELUTING CORONARY STENT SYSTEM

MDR report key: 1023698 · Received April 1, 2008

Report

Report Number
2134265-2008-00945
Event Type
Malfunction
Date Received
April 1, 2008
Date of Event
March 12, 2008
Report Date
March 12, 2008
Manufacturer
BOSTON SCIENTIFIC
Product Code
NIQ
PMA / PMN Number
P030025
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A CORONARY DRUG ELUTING STENTING TREATMENT PROCEDURE, A BALLOON PINHOLE RUPTURE AND DEPLOYMENT DIFFICULTY OCCURRED. THE PHYSICIAN ATTEMPTED TO PLACE A TAXUS EXPRESS2 3.0X24CM DRUG ELUTING STENT TO THE 90% STENOSED LESION LOCATED IN THE CALCIFIED, EXTREMELY TORTUOUS DISTAL RIGHT CORONARY ARTERY (RCA), BUT WAS UNABLE TO CROSS THE LESION. AN IVUS CATHETER WAS ALSO UNABLE TO CROSS THE LESION. THE LESION WAS THEN PRE-DILATED SEVERAL TIMES WITH A NON BSC 2.5X15MM BALLOON. THE TAXUS 3.0X24MM STENT WAS ABLE TO CROSS THE LESION. AN ATTEMPT WAS MADE TO DILATE THE STENT BALLOON, BUT IT WOULD ONLY DILATE UP TO 5 ATMS. THE DISTAL SIDE OF THE BALLOON WAS "PINHOLE RUPTURED." UPON WITHDRAWAL, THE STENT DELIVERY SYSTEM CAUGHT ON "SOMETHING" IN THE MID RCA. THERE WAS A LESION LOCATED IN THE MID RCA, SO THE 3.0X24MM TAXUS EXPRESS2 DRUG ELUTING STENT WAS IMPLANTED THERE AT A LOW PRESSURE. IVUS WAS THEN PERFORMED AND IT WAS NOTED THAT PART OF THE STENT WAS DILATED INCOMPLETELY. POST DILATATION WAS THEN PERFORMED WITH THE NON BSC BALLOON. IVUS CONFIRMED COMPLETE STENT WAS THEN IMPLANTED DISTAL TO THE 3.0X24MM STENT AND A TAXUS 3.5X28MM DRUG ELUTING STENT WAS IMPLANTED PROXIMAL TO THE 3.0X24MM STENT. NO PT COMPLICATIONS WERE REPORTED. PT STATUS POST PROCEDURE IS NOTED AS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TAXUS EXPRESS2 PACLITAXEL-ELUTING CORONARY STENT SYSTEM NIQ STENT, CORONARY, DRUG-ELUTING NIQ BOSTON SCIENTIFIC H7493897024300

Patients

Seq Age Sex Outcome Treatment
1 RYUJIN BALLOON