FDA Adverse Event Injury Summary report: N

ACS HI-TORQUE BALANCE MIDDLEWEIGHT GUIDE WIRE

MDR report key: 1023244 · Received March 31, 2008

Report

Report Number
2024168-2008-00240
Event Type
Injury
Date Received
March 31, 2008
Date of Event
March 5, 2008
Report Date
March 6, 2008
Manufacturer
ABBOTT VASCULAR-CARDIAC THERAPIES
Product Code
DQX
PMA / PMN Number
K013833
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT PERFORMANCE ENGINEERING REVIEWED THE INCIDENT INFO. SIMILAR REPORTS ON RETURNED DEVICES TYPICALLY SHOWS THAT GUIDE WIRE TIP DAMAGE MAY HAPPEN WHEN THE CORE IS SUBJECTED TO TENSILE OR TORSIONAL LOADS BEYOND ITS DESIGN LIMITS CAUSING THE DISTAL TIP TO DETACH. TYPICALLY, THE FRACTURE SITE MORPHOLOGY SHOWS THE CORE WAS EXPOSED TO FORCES CONSISTENT WITH THOSE APPLIED THROUGH PULLING, TORQUING, AND MANIPULATION. A GUIDE WIRE BEING OVER PULLED / OVER TORQUED WOULD REQUIRE THE TIP TO BE TRAPPED WITHIN THE ANATOMY OR ANOTHER DEVICE IN ORDER TO CREATE THE REQUIRED FORCES TO DAMAGE THE GUIDE WIRE. THE GUIDE WIRE USED IN THIS CASE WAS NOT RETURNED; THEREFORE, A DEFINITIVE ROOT CAUSE OF THE REPORTED ISSUE CANNOT BE DETERMINED.

Description of Event or Problem · 1

REPORTING STATUS: SERIOUS INJURY - REQUIRED INTERVENTION/HOSPITALIZATION. REPORTING RATIONALE: GUIDE WIRE SEPARATION REQUIRING SURGICAL INTERVENTION. DEVICE ISSUE: GUIDE WIRE SEPARATION. IT WAS REPORTED THAT DURING PLACEMENT OF ANOTHER COMPANY'S STENT, RESISTANCE WAS MET WITH THE STENT. THE GUIDE WIRE SEPARATED IN THE PT. THE DEVICES WERE THEN REMOVED AS A UNIT; HOWEVER, AFTER REMOVAL IT APPEARED AS THOUGH SOME OF COILS WERE LEFT IN THE PT. THE PT WAS THEN SENT TO SURGERY FOR REMOVAL. ALL PORTIONS OF THE DEVICE WERE ABLE TO BE REMOVED AND THERE WAS NO PORTION LEFT IN THE PT. PT WAS DOING WELL POST SURGERY. NO ADD'L EVENT OR PT INFO IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACS HI-TORQUE BALANCE MIDDLEWEIGHT GUIDE WIRE DQX ABBOTT VASCULAR-CARDIAC THERAPIES NA UNK

Patients

Seq Age Sex Outcome Treatment
1 UNK Hospitalization| R