COZMO INSULIN PUMP
Report
- Report Number
- 2183502-2008-00067
- Event Type
- Injury
- Date Received
- March 31, 2008
- Date of Event
- February 27, 2008
- Report Date
- March 29, 2008
- Manufacturer
- SMITHS MEDICAL MD (FORMERLY DELTEC, INC.)
- Product Code
- LZG
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- LAY USER/PATIENT
Narratives
EVAL SUMMARY: THE SUSPECT DEVICE WAS RETURNED AND EVALUATED. DELIVERY AND ACCURACY TESTS WERE PERFORMED, THE DEVICE WAS FOUND TO PASS ALL DELIVERY AND ACCURACY TESTS. THE PUMP HISTORY WAS DOWNLOADED AND ANALYZED NO ANOMALIES WERE FOUND. NO OPERATIONAL OR FUNCTIONAL FAILURE WAS DETECTED AND THE PRODUCT WAS WITHIN SPECIFICATION.
INFO WAS RECEIVED THAT REPORTED A PT WAS HOSPITALIZED IN 2008 DUE TO AN INCIDENT OF HYPERGLYCEMIA. THE REPORTER STATED SHE AWOKE IN THE A.M. ON THE SAME DAY AND HER BLOOD GLUCOSE WAS 109 MG/DL. SHE DID NOT EAT OR DRINK AND WENT TO WORK, WHERE SHE TESTED AND WAS 320 MG/DL. TWO HRS LATER, SHE WAS 490 MG/DL. SHE WENT TO THE ER AND WAS TREATED WITH 2 LITERS OF FLUID AND 20 UNITS OF INSULIN. THE INSULIN PUMP WILL BE RETURNED FOR EVAL; TO ENSURE THAT IT IS FUNCTIONING CORRECTLY AND DID NOT CONTRIBUTE TO THE REPORTED INCIDENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COZMO INSULIN PUMP | INSULIN PUMP | LZG | SMITHS MEDICAL MD (FORMERLY DELTEC, INC.) | 1700 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |