COLLEAGUE 3 VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2007-01668
- Event Type
- Malfunction
- Date Received
- April 4, 2008
- Date of Event
- December 22, 2006
- Report Date
- December 22, 2006
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Removal / Correction Number
- 6000001-7/20/05-015-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 16 2007. EVALUATION SUMMARY:THE REPORTED CONDITION OF FAILURE CODE 703 ON CHANNELS B AND C WAS CONFIRMED IN THE EVENT HISTORY BY THE BAXTER REPAIR TECHNICIAN. INSPECTION OF THE DEVICE REVEALED THE FAILURE WAS CAUSED BY A DEFECTIVE UIM PCB (USER INTERFACE MECHANISM PRINTED CIRCUIT BOARD), WHICH WAS REPLACED. THE DEVICE WAS TESTED BY THE REPAIR TECHNICIAN AND RETURNED TO SERVICE. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS CURRENTLY BEING INVESTIGATED UNDER CAPA.
THE DEVICE WAS RETURNED FOR SERVICE. DURING SERVICE TESTING, THE BAXTER TECHNICIAN REPORTED AN INFUSION PUMP WITH A FAILURE CODE 703 ON CHANNELS B AND C. PER THE SERVICE SHOP, THE HOSPITAL REPRESENTATIVE STATED THAT THERE HAVE BEEN NO REPORTS OF PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS KNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE 3 VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE. LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |