FDA Adverse Event
Malfunction
Summary report: N
COLLEAGUE 3 VOLUMETRIC INFUSION PUMP
MDR report key: 1023178
·
Received April 4, 2008
Report
- Report Number
- 6000001-2007-00931
- Event Type
- Malfunction
- Date Received
- April 4, 2008
- Date of Event
- December 1, 2006
- Report Date
- December 11, 2006
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
EVALUATION SUMMARY: THE DEVICE EVALUATION WAS COMPLETED AND FAILURE CODE 500 CONFIRMED (KEYPAD TEST FAILED #8) DUE TO A DEFECTIVE FRONT BEZEL. SERVICE INSTALLED NEW FRONT BEZEL AND TESTED THE DEVICE. A REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT AND THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA.
Description of Event or Problem · 1
THE FACILITY REPRESENTATIVE REPORTED A PUMP WITH FAILURE CODE 500 (KEYPAD TEST FAILED FOR #8). THE CONDITION WAS DISCOVERED DURING BIO-MED TESTING. ACCORDING TO THE HOSPITAL REPRESENTATIVE, THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE 3 VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE. LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |