FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX VOLUMETRIC INFUSION PUMP

MDR report key: 1023173 · Received April 4, 2008

Report

Report Number
6000001-2007-04072
Event Type
Malfunction
Date Received
April 4, 2008
Date of Event
January 13, 2007
Report Date
January 13, 2007
Manufacturer
BAXTER HEALTHCARE PTE LTD
Product Code
FRN
Removal / Correction Number
CONTINUED ON H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE CONDITION OF DEPLETED MAIN BATTERIES WAS CONFIRMED. THE MAIN BATTERIES WERE 4 DISCHARGES BELOW ALARM THRESHOLD. THE MAIN BATTERIES WERE REPLACED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS CURRENTLY BEING INVESTIGATED.

Description of Event or Problem · 1

THE FACILITY RETURNED THE DEVICE FOR SERVICE. DURING SERVICE, THE BAXTER REPAIR TECHNICIAN NOTED DEPLETED BATTERIES WITH 4 DISCHARGES BELOW ALARM THRESHOLD. THE HOSPITAL REPRESENTATIVE STATED THAT THERE HAVE BEEN NO REPORTS OF PATIENT INCIDENT INVOLVING THIS PUMP. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CX VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE LTD NA NA

Patients

Seq Age Sex Outcome Treatment
1