FDA Adverse Event
Malfunction
Summary report: N
COLLEAGUE CX VOLUMETRIC INFUSION PUMP
MDR report key: 1023173
·
Received April 4, 2008
Report
- Report Number
- 6000001-2007-04072
- Event Type
- Malfunction
- Date Received
- April 4, 2008
- Date of Event
- January 13, 2007
- Report Date
- January 13, 2007
- Manufacturer
- BAXTER HEALTHCARE PTE LTD
- Product Code
- FRN
- Removal / Correction Number
- CONTINUED ON H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
EVALUATION SUMMARY: THE CONDITION OF DEPLETED MAIN BATTERIES WAS CONFIRMED. THE MAIN BATTERIES WERE 4 DISCHARGES BELOW ALARM THRESHOLD. THE MAIN BATTERIES WERE REPLACED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS CURRENTLY BEING INVESTIGATED.
Description of Event or Problem · 1
THE FACILITY RETURNED THE DEVICE FOR SERVICE. DURING SERVICE, THE BAXTER REPAIR TECHNICIAN NOTED DEPLETED BATTERIES WITH 4 DISCHARGES BELOW ALARM THRESHOLD. THE HOSPITAL REPRESENTATIVE STATED THAT THERE HAVE BEEN NO REPORTS OF PATIENT INCIDENT INVOLVING THIS PUMP. NO ADDITIONAL INFORMATION IS AVAILABLE.