FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX VOLUMETRIC INFUSION PUMP

MDR report key: 1023132 · Received April 3, 2008

Report

Report Number
6000001-2007-03245
Event Type
Malfunction
Date Received
April 3, 2008
Date of Event
January 8, 2007
Report Date
January 8, 2007
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Removal / Correction Number
CONTINUED ON H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE CONDITION OF DAMAGED BATTERIES WITH 2 BATTERY DISCHARGES BELOW THE ALARM THRESHOLD WAS CONFIRMED. THE PUMP'S BATTERIES WERE REPLACED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED.

Description of Event or Problem · 1

DURING PRODUCT EVALUATION, THE PUMP'S BATTERIES WERE FOUND TO BE DAMAGED WITH 2 BATTERY DISCHARGES BELOW THE ALARM THRESHOLD. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CX VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1