FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX VOLUMETRIC INFUSION PUMP

MDR report key: 1023126 · Received April 3, 2008

Report

Report Number
6000001-2007-01894
Event Type
Malfunction
Date Received
April 3, 2008
Date of Event
December 27, 2006
Report Date
December 27, 2006
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Removal / Correction Number
CONTINUED ON H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE CONDITION OF DEPLETED BATTERIES WAS CONFIRMED. INSPECTION OF THE DEVICE FOUND THAT THE PUMP'S BATTERIES WERE POTENTIALLY DAMAGED. THE BATTERIES WERE REPLACED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS CURRENTLY BEING INVESTIGATED UNDER CAPA, WHICH WAS OPENED ON APRIL 1, 2005.

Description of Event or Problem · 1

DURING PRODUCT EVALUATION THE PUMP'S BATTERIES WERE FOUND TO BE DEPLETED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CX VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1