COLLEAGUE 3CX VOLUMETRIC COLLEAGUE VOLUMETRIC PUMP
Report
- Report Number
- 6000001-2007-00759
- Event Type
- Malfunction
- Date Received
- April 3, 2008
- Date of Event
- December 12, 2006
- Report Date
- December 12, 2006
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Removal / Correction Number
- CONTINUED ON H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 08 2007. EVALUATION SUMMARY:THE CONDITION OF FAILURE CODE 570:320:844 WAS CONFIRMED IN THE PUMP'S EVENT HISTORY FILE DURING PRODUCT EVALUATION. INSPECTION OF THE DEVICE FOUND THE PUMP'S MAIN BATTERIES WERE FOUR DISCHARGES BELOW THEIR ALARM THRESHOLD. THE PUMP'S MAIN BATTERIES WERE DAMAGED AND THEREFORE THE BATTERIES WERE REPLACED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS CURRENTLY BEING INVESTIGATED UNDER CAPA.
DURING PRODUCT EVALUATION, FAILURE CODE 570:320:844 WAS IDENTIFIED IN THE PUMP'S EVENT HISTORY FILE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE 3CX VOLUMETRIC COLLEAGUE VOLUMETRIC PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE. LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |