FDA Adverse Event Malfunction Summary report: N

COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP

MDR report key: 1023050 · Received April 3, 2008

Report

Report Number
6000001-2007-02604
Event Type
Malfunction
Date Received
April 3, 2008
Date of Event
September 1, 2006
Report Date
September 1, 2006
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Removal / Correction Number
6000001-7/20/05-015-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE DEVICE WAS EVALUATED AT A BAXTER SERVICE CENTER. THE REPORTED CONDITION OF FAILURE CODE 703 WAS CONFIRMED. THE USER INTERFACE MODULE PRINTED CIRCUIT BOARD WAS FOUND TO BE OUT OF SPECIFICATION WHICH CAUSED THE FAILURE TO OCCUR. THE CIRCUIT BOARD WAS REPLACED AND THE DEVICE WAS RETURNED TO THE CUSTOMER FULLY OPERATIONAL. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED FROM CUSTOMER FOR SERVICE. DURING SERVICE BY BAXTER TECHNICIAN, THE DEVICE WAS FOUND TO HAVE FAILURE CODE 703 IN THE EVENT HISTORY. CUSTOMER REPORTED THERE WERE NO PATIENT INJURIES OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. NA NA