FDA Adverse Event
Malfunction
Summary report: N
COLLEAGUE VOLUMETRIC INFUSION PUMP
MDR report key: 1023034
·
Received April 3, 2008
Report
- Report Number
- 6000001-2007-02513
- Event Type
- Malfunction
- Date Received
- April 3, 2008
- Date of Event
- September 6, 2006
- Report Date
- September 6, 2006
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
EVALUATION SUMMARY: THE INFUSION PUMP WAS TESTED FOR FLOW ACCURACY AT 100ML/HR, THE PUMP FAILED THE ACCURACY TEST WITH A FLOW VALUE BELOW THE DESIRED AMOUNT. THE SPECIFICATION LIMIT FOR THIS PUMP IS +/-5%. THE PUMPHEAD WAS REPLACED AND THE DEVICE WAS RETURNED TO THE CUSTOMER FULLY OPERATIONAL.
Description of Event or Problem · 1
THE DEVICE WAS RETURNED FROM CUSTOMER FOR SERVICE. DURING SERVICE BY BAXTER TECHNICIAN, THE DEVICE FAILED ACCURACY TEST WITH UNDERDELIVERY. CUSTOMER REPORTED THERE WERE NO PATIENT INJURIES OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE. LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |