FDA Adverse Event Malfunction Summary report: N

MICRA

MDR report key: 10229005 · Received July 3, 2020

Report

Report Number
9612164-2020-02405
Event Type
Malfunction
Date Received
July 3, 2020
Date of Event
June 30, 2020
Report Date
August 6, 2020
Manufacturer
MEDTRONIC IRELAND
Product Code
PNJ
PMA / PMN Number
P150033
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

PRODUCT ID# MC1VR01-DELSYS D10: YES, RETURN DATE: 09-JUL-2020 H3: YES H6: FDA METHOD CODE(S): 10 H6: FDA RESULTS CODE(S): 3243, 180, 3038 H6: FDA CONCLUSION CODE(S): 4315 PRODUCT EVENT SUMMARY: THE FULL DELIVERY SYSTEM WAS RETURNED AND ANALYZED. THE ANALYSIS INDICATED THAT THE DELIVERY SYSTEM OUTER SHAFT WAS KINKED/BUCKLED. THE ARTICULATION OF THE DELIVERY SYSTEM WAS OUT OF PLANE. THE ARTICULATION CURVE OF THE DELIVERY SYSTEM DID NOT MEET SPECIFICATION. THE LUMEN OF THE DELIVERY SYSTEM WAS DAMAGED. THE DELIVERY SYSTEM INNER SHAFT WAS MECHANICALLY KINKED/BENT. BLOOD WAS OBSERVED IN THE LUMEN OF THE DELIVERY SYSTEM. BLOOD WAS OBSERVED ON THE DEVICE CUP OF THE DELIVERY SYSTEM. BLOOD WAS OBSERVED ON THE RECAPTURE CONE OF THE DELIVERY SYSTEM. VISUAL ANALYSIS OF THE DELIVERY SYSTEM INDICATED DAMAGE DURING USE. THE ANALYST NOTED THE FULL DELIVERY SYSTEM WAS RETURNED WITH THE DEVICE INTACT IN THE DEVICE CUP. THE OUTER SHAFT IS KINK/BUCKLED AT 5 CM FROM THE DISTAL END OF THE DELIVERY SYSTEM. THE INNER SHAFT IS KINKED AT 1.5 CM AND 2.8 CM FROM THE DISTAL END OF THE RECAPTURE CONE. THERE IS BLOOD ON THE OUTER SHAFT LOCATED AT THE DISTAL END OF THE STABILITY MEMBER. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. RETURNED PRODUCT ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

OTHER RELEVANT DEVICE(S) ARE: MICRA, MODEL #: MC1VR01-DELSYS / EXPIRATION DATE: 2021-04-28 / SERIAL#: (B)(4), UDI #: (B)(4), DEV RTN TO MFR? NO, MFG DATE: 2019-10-28, LABELED FOR SINGLE USE: YES, (B)(4). IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT, DURING THE IMPLANT PROCEDURE, THE LEADLESS IMPLANTABLE PULSE GENERATOR (IPG), AFTER FIVE ATTEMPTS COULD NOT BE RE POSITIONED AND THE TIP OF THE CATHETER 'WAS SOFT'. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
693333 MICRA LEADLESS PACEMAKER PNJ MEDTRONIC IRELAND MC1VR01

Patients

Seq Age Sex Outcome Treatment
1