IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 10MM
Report
- Report Number
- 0002023141-2020-00999
- Event Type
- Injury
- Date Received
- July 3, 2020
- Date of Event
- June 4, 2020
- Report Date
- July 27, 2020
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- PMA / PMN Number
- K061410
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
ONE IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 10MM (TSVWB10) WAS RETURNED FOR INVESTIGATION. VISUAL EVALUATION OF THE AS RETURNED PRODUCT IDENTIFIED MINIMAL SIGNS OF WEAR AND DEBRIS ABOUT THE THREADS AND DRIVE FEATURE. IT IS INDICATED THAT THE PATIENT HAS A HISTORY OF BRUXISM AND CLENCHING. THE REPORTED PRODUCT WAS LOCATED ON TOOTH # 3 AND USED FOR APPROXIMATELY 6 MONTHS. THE REPORTED EVENT COULD NOT BE RECREATED DUE TO THE NATURE OF THE DENTAL DEVICE AND EVENT (MEDICAL CONDITIONS). THE CUSTOMER HAS NOT PROVIDED ADDITIONAL PICTURES OR X-RAY IMAGES OF THE PRODUCT. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1230746. IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (1230746) FOR SIMILAR EVENT AND NO OTHER COMPLAINT WAS IDENTIFIED. MAY POST MARKET TRENDING WAS REVIEWED AND THERE WERE NO ACTIONABLE EVENTS OR CORRECTIVE ACTIONS FOR THE REPORTED EVENT (SINUS PERFORATION) OR PRODUCT (TSVWB10). THEREFORE, BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION HAS NOT OCCURRED AND THE REPORTED EVENT WAS NON-VERIFIABLE. A DEFINITIVE CAUSE COULD NOT BE IDENTIFIED. HOWEVER, BASED ON THE INVESTIGATION AND RISK FILE REVIEW, THE POTENTIAL CAUSES FOR THE REPORTED EVENT IS CLINICIAN ERROR IN CASE PLANNING.
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
ZIMMER BIOMET COMPLAINT (B)(4). PATIENT WEIGHT: NOT PROVIDED. ADDITIONAL PMA/ 510(K) NUMBER: K011028, K013227.
IT WAS REPORTED THAT IMPLANT (TSVWB10) FAIL TO INTEGRATE, PERFORATED THE MAXILLARY SINUS AND LATER REMOVED. INFLAMMATION WAS ALSO REPORTED. TOOTH LOCATION 3.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 691598 | IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 10MM | DENTAL IMPLANT | DZE | ZIMMER DENTAL | 1230746 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR | Required Intervention |