FDA Adverse Event Malfunction Summary report: N

ALERE AFINION AS100 ANALYZER

MDR report key: 10228327 · Received July 3, 2020

Report

Report Number
3003045237-2020-00006
Event Type
Malfunction
Date Received
July 3, 2020
Date of Event
June 4, 2020
Report Date
July 2, 2020
Manufacturer
ABBOTT DIAGNOSTICS TECHNOLOGIES AS
Product Code
JQT
PMA / PMN Number
K151809
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A INVESTIGATION IS ONGOING FOR AFINION ANALYZERS RELATED TO SIMILAR INCIDENTS AND THE INVESTIGATION TO DATE HAS CONCLUDED THAT ELECTRIC SHOCK HAS NOT BEEN DEMONSTRATED WITH THE ANALYZERS. RATHER, STATIC DISCHARGE MAY BE POSSIBLE AND IS MORE CONSISTENT WITH THE DESCRIPTION OF THE INCIDENT. PER THE HEALTH HAZARD EVALUATION STATIC DISCHARGE DOES NOT POSE A RISK TO HARM WHEREAS BREACH OF POWER SUPPLY COULD HAVE THE POTENTIAL TO RESULT IN BURN INJURY. AS THE COMPLAINT IS NOT CLEAR IF THIS COULD BE A POTENTIAL BREACH, AN MDR IS FILED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE AFINION AS100 ANALYZER SPARKED WHEN PLUGGED INTO THE WALL. THERE WAS NO CHARRING OR BURNING. THE CUSTOMER DISCONNECTED THE INSTRUMENT AFTER THE INCIDENT AND THE INSTRUMENT HAS NOT BEEN USED SINCE. THE CUSTOMER DESCRIBED THAT SHE FELT AN ELECTRIC SHOCK IN HER ARM. NO INJURY WAS REPORTED. THE CUSTOMER DID NOT PROVIDE FURTHER DETAILS ON THE SPARKING AND IT COULD BE CAUSED BY ELECTROSTATIC DISCHARGE.