ALERE AFINION AS100 ANALYZER
Report
- Report Number
- 3003045237-2020-00006
- Event Type
- Malfunction
- Date Received
- July 3, 2020
- Date of Event
- June 4, 2020
- Report Date
- July 2, 2020
- Manufacturer
- ABBOTT DIAGNOSTICS TECHNOLOGIES AS
- Product Code
- JQT
- PMA / PMN Number
- K151809
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
A INVESTIGATION IS ONGOING FOR AFINION ANALYZERS RELATED TO SIMILAR INCIDENTS AND THE INVESTIGATION TO DATE HAS CONCLUDED THAT ELECTRIC SHOCK HAS NOT BEEN DEMONSTRATED WITH THE ANALYZERS. RATHER, STATIC DISCHARGE MAY BE POSSIBLE AND IS MORE CONSISTENT WITH THE DESCRIPTION OF THE INCIDENT. PER THE HEALTH HAZARD EVALUATION STATIC DISCHARGE DOES NOT POSE A RISK TO HARM WHEREAS BREACH OF POWER SUPPLY COULD HAVE THE POTENTIAL TO RESULT IN BURN INJURY. AS THE COMPLAINT IS NOT CLEAR IF THIS COULD BE A POTENTIAL BREACH, AN MDR IS FILED.
THE CUSTOMER REPORTED THAT THE AFINION AS100 ANALYZER SPARKED WHEN PLUGGED INTO THE WALL. THERE WAS NO CHARRING OR BURNING. THE CUSTOMER DISCONNECTED THE INSTRUMENT AFTER THE INCIDENT AND THE INSTRUMENT HAS NOT BEEN USED SINCE. THE CUSTOMER DESCRIBED THAT SHE FELT AN ELECTRIC SHOCK IN HER ARM. NO INJURY WAS REPORTED. THE CUSTOMER DID NOT PROVIDE FURTHER DETAILS ON THE SPARKING AND IT COULD BE CAUSED BY ELECTROSTATIC DISCHARGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 694630 | ALERE AFINION AS100 ANALYZER | ANALYZER | JQT | ABBOTT DIAGNOSTICS TECHNOLOGIES AS | 1115175 | AS0057766 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |