FDA Adverse Event Malfunction Summary report: N

LUPINE DRILL GUIDE

MDR report key: 1022484 · Received March 31, 2008

Report

Report Number
1221934-2008-00165
Event Type
Malfunction
Date Received
March 31, 2008
Date of Event
March 24, 2008
Report Date
March 24, 2008
Manufacturer
DEPUY MITEK
Product Code
HRX
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

WE BELIEVE THAT THIS TYPE OF EVENT IS MOST LIKELY TECHNIQUE RELATED, EXTENSIVE LAB TESTING HAS SHOWN THAT UNDER NORMAL CONDITIONS, THE REPORTED EVENT MODE COULD NOT BE DUPLICATED, HOWEVER, UNDER EXTREME CONDITIONS, THAT IS WHEN THE DRILL GUIDE WAS BENT DURING DRILLING WITH A DRILL BIT, THIS CAUSED THE DRILL BIT TO RUB AGAINST INNER SURFACE OF THE BENT DRILL GUIDE CAUSING SOME METAL TO SHAVE OFF OF THE DRILL GUIDE, THE REPORTED CONDITION WAS THEN DUPLICATED. THIS PHENOMENA HAS BEEN THE SUBJECT OF A LONG AND CONSIDERABLE STUDY. AT THIS POINT IN TIME, THE ENGINEERING DESIGN AND THE QUALITY ENGINEERING TEAMS ARE IN THE MIDST OF DEVELOPING SOME DESIGN CHANGED TO SUBVERT AND OR GREATLY REDUCE THIS TYPE OF EVENT. MITEK WILL CONTINUE TO TRACK ANY RELATED COMPLAINTS WITHIN THIS DEVICE FAMILY AS A MEANS OF MONITORING THE EXTENT WITH WHICH THIS COMPLAINT IS OBSERVED IN THE FIELD. WHEN AND IF THE COMPLAINT DEVICE IS RECEIVED HERE AT MITEK, IT WILL BE DISCARDED AS THE REPORTED PHENOMENON IS ACKNOWLEDGED AND IS IN THE PROCESS OF BEING ADDRESSED.

Description of Event or Problem · 1

OUR REP IS REPORTING THAT DURING AN ARTHROSCOPIC SHOULDER PROCEDURE, THE SURGEON OBSERVED METAL SHAVINGS EMANATING FROM THE LUPINE DRILL GUIDE FALLING INTO THE PT'S JOINT SPACE. ALL OF THE DEBRIS WAS REMOVED FROM THE BODY AND THE PROCEDURE WAS CONCLUDED SUCCESSFULLY WITHOUT FURTHER INCIDENT OR HARM TO THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LUPINE DRILL GUIDE ARTHROSCOPIC INSTRUMENT HRX DEPUY MITEK 213817 UNK

Patients

Seq Age Sex Outcome Treatment
1 UNK