LUPINE DRILL GUIDE
Report
- Report Number
- 1221934-2008-00164
- Event Type
- Malfunction
- Date Received
- March 31, 2008
- Date of Event
- March 19, 2008
- Report Date
- March 20, 2008
- Manufacturer
- DEPUY MITEK
- Product Code
- HRX
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
WE BELIEVE THAT THIS TYPE OF EVENT IS MOST LIKELY TECHNIQUE RELATED, EXTENSIVE LAB TESTING HAS SHOWN THAT UNDER NORMAL CONDITIONS THE REPORTED EVENT MODE COULD NOT BE DUPLICATED, HOWEVER, UNDER EXTREME CONDITIONS, THAT IS WHEN THE DRILL GUIDE WAS BENT DURING DRILLING WITH A DRILL BIT, THIS CAUSED THE DRILL BIT TO RUB AGAINST INNER SURFACE OF THE BENT DRILL GUIDE CAUSING SOME METAL TO SHAVE OFF OF THE DRILL GUIDE, THE REPORTED CONDITION WAS THEN DUPLICATED. THIS PHENOMENA HAS BEEN THE SUBJECT OF A LONG AND CONSIDERABLE STUDY. AT THIS POINT IN TIME, THE ENGINEERING DESIGN AND THE QUALITY ENGINEERING TEAMS ARE IN THE MIDST OF DEVELOPING SOME DESIGN CHANGED TO SUBVERT AND OR GREATLY REDUCE THIS TYPE OF EVENT. MITEK WILL CONTINUE TO TRACK ANY RELATED COMPLAINTS WITHIN THIS DEVICE FAMILY AS A MEANS OF MONITORING THE EXTENT WITH WHICH THIS COMPLAINT IS OBSERVED IN THE FIELD. WHEN AND IF THE COMPLAINT DEVICE IS RECEIVED HERE AT MITEK, IT WILL BE DISCARDED AS THE REPORTED PHENOMENON IS ACKNOWLEDGED AND IS IN THE PROCESS OF BEING ADDRESSED.
OUR REP IS REPORTING THAT DURING AN ARTHROSCOPIC SHOULDER PROCEDURE, THE SURGEON OBSERVED METAL SHAVINGS EMANATING FROM THE LUPINE DRILL GUIDES FALLING INTO THE PT'S JOINT SPACE. ALL OF THE DEBRIS WAS REMOVED FROM THE BODY AND THE PROCEDURE WAS CONCLUDED SUCCESSFULLY WITHOUT FURTHER INCIDENT OR HARM TO THE PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LUPINE DRILL GUIDE | ARTHROSCOPIC INSTRUMENT | HRX | DEPUY MITEK | 213816 | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |