FDA Adverse Event Malfunction Summary report: N

I-STAT 6+ CARTRIDGE

MDR report key: 1022465 · Received March 31, 2008

Report

Report Number
2245578-2008-00010
Event Type
Malfunction
Date Received
March 31, 2008
Date of Event
July 12, 2007
Report Date
March 25, 2008
Manufacturer
ABBOTT POINT OF CARE INC.
Product Code
JJE
Removal / Correction Number
Z-0370/375-08
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THERE IS A COMPONENT ON THE COVER OF THE CARTRIDGE CALLED THE SNAP CLOSURE. THE SNAP CLOSURE CREATES AN AIRTIGHT SEAL NECESSARY FOR PROPER FLUID MOVEMENT WITHIN THE CARTRIDGE. THE CLOSURE ENSURES THE CALIBRANT AND SAMPLE REMAIN CONTAINED WITHIN THE CARTRIDGE DURING THE TESTING CYCLE AND SUBSEQUENT DISPOSAL. THE USER IS INSTRUCTED TO FOLD THE SNAP CLOSURE OVER THE SAMPLE WELL AND TO PRESS THE ROUNDED END OF THE CLOSURE UNTIL IT SNAPS INTO PLACE AFTER A LOADING SAMPLE. THE SNAP CLOSURE REMAINS IN PLACE BY A LATCH HOOK. IN THE CASE OF THIS PRODUCT ISSUE, THE SNAP CLOSURE WAS HARD TO CLOSE AND/OR DID NOT REMAIN CLOSED. AN INVESTIGATION IDENTIFIED AN ASSIGNABLE CAUSE FOR THE ISSUE; A NON-OPTIMAL DIMENSION OF THE SNAP CLOSURE MECHANISM. IT WAS FOUND THAT WHEN THE DIMENSION OF THE LATCH HOOK IS CLOSER TO THE LOWER END OF THE SPECIFICATION MORE FORCE IS REQUIRED TO ENGAGE THE HOOK WITH THE UNDERSIDE OF THE COVER THAN WHEN IT IS TOWARD THE UPPER END OF THE SPECIFICATION. CONSEQUENTLY, THE PROBABILITY IS HIGHER THAT A FORCE IS GENERATED UPON CLOSURE OF THE LATCH THAT IS LARGE ENOUGH TO DEFORM THE PLASTIC WHERE THE HOOK MUST ENGAGE WITH THE UNDERSIDE OF THE COVER. ONCE THE PLASTIC IS DEFORMED, THE SNAP CLOSURE CANNOT CLOSE OR WILL NOT REMAIN CLOSED.

Description of Event or Problem · 1

A CUSTOMER STATED THE SNAP CLOSURE ON AN I-STAT 6+ CARTRIDGE IS DIFFICULT TO CLOSE. THERE WAS NO REPORT OF ANY INJURY ASSOCIATED WITH THE COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 I-STAT 6+ CARTRIDGE IVD JJE ABBOTT POINT OF CARE INC. NA D07116

Patients

Seq Age Sex Outcome Treatment
1