HDF3500 AND PAD-PAK
Report
- Report Number
- 3004123209-2020-00168
- Event Type
- Malfunction
- Date Received
- July 2, 2020
- Date of Event
- July 1, 2018
- Report Date
- July 2, 2020
- Manufacturer
- HEARTSINE TECHNOLOGIES LTD
- Product Code
- MKJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JP
- Reporter Occupation
- OTHER
Narratives
THE DEVICE HISTORY RECORDS FOR THE HDF-3500 DEVICE WAS REVIEWED AND THIS CONFIRMED THAT ALL MANUFACTURING AND QUALITY CHECKS AND TEST HAD BEEN SUCCESSFULLY COMPLETED. THE HDF-3500 PASSED ¿OUT QAT FROM HEARTSINE TECHNOLOGIES ON THE 10TH JUNE 2018.THE DEVICE WAS RETURNED FOR ISSUING A LOW BATTERY PROMPT DURING OMRON GOODS INWARD TESTING (H045-014-005), WHILE CONNECTED TO A TEST POWER SUPPLY. DESPITE BEING UNABLE TO REPEAT THE FAULT DURING THE INVESTIGATION, THERE WAS AN OBSERVABLE DEFECTIVE DETECTED WITH THE -VE BATTERY TERMINAL POGO PIN. THE -VE POGO PIN HAD FAILED TO SPRING INTO THE NORMAL POSITION AND DID NOT PRESENT A SIGNIFICANT RESISTANCE WHEN PRESSED. THIS DEFECT WOULD ACCOUNT FOR THE LOW BATTERY WARNING, DUE TO AN INTERMITTENT BATTERY TERMINAL CONNECTION WITH THE INSTALLED TEST PAD-PAK INTERFACE. THE POGO PINS ARE CHECKED DURING HDF-3500 LOWER CASE ASSEMBLY (H006-006-187). THE DEVICE ALSO SUCCESSFULLY PASSED THE HDF-3500 FINAL UNIT TEST (H045-014-004) WHILE AT HEARTSINE, THEREFORE THE FAULT WOULD HAVE EITHER HAD TO HAVE OCCURRED AFTER THIS TIME OR WAS INTERMITTENT IN NATURE. IT IS A POLICY OF HEARTSINE TO NOT REFURBISH DEVICES THAT HAVE BEEN RETURNED FROM THE FIELD, THEREFORE THIS DEVICE SHALL BE RETAINED AND REPLACED WITH A NEW HDF-3500. .
DEVICE SERVICE REQUIRED PROMPT. THERE WAS NO PATIENT INVOLVED IN THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 687317 | HDF3500 AND PAD-PAK | AUTOMATED EXTERNAL DEFIBRILLATOR | MKJ | HEARTSINE TECHNOLOGIES LTD |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |