FDA Adverse Event Injury Summary report: N

CORAL SPINAL SYSTEM

MDR report key: 10222277 · Received July 1, 2020

Report

Report Number
3012120772-2020-00047
Event Type
Injury
Date Received
July 1, 2020
Report Date
July 1, 2020
Manufacturer
SEASPINE INC.
Product Code
NKB
PMA / PMN Number
K120047
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A REVISION SURGERY OCCURRED ON (B)(6) 2017 AS A RESULT OF THE PROMIXIMAL JUNCTIONAL FAILURE WITH PROXIMAL SCREW PULL OUT. NO IMPLANTS WERE RETURNED FOR ANALYSIS AND THE PATIENT WAS DISCHARGED WITH STABLE APPEARANCE TO CONSTRUCT, AND NEUROLOGICALLY INTACT. REVIEW OF LABELING: POSTOPERATIVE WARNINGS SURGEONS SHOULD ADVISE PATIENTS REGARDING THE RISKS OF SURGERY AND THE IMPORTANCE OF POST-OPERATIVE COMPLIANCE. THE PATIENT SHOULD BE ADVISED TO LIMIT AND RESTRICT PHYSICAL ACTIVITIES, ESPECIALLY LIFTING AND TWISTING MOTIONS AND ANY TYPE OF SPORT PARTICIPATION. THE PATIENT SHOULD BE ADVISED THAT IMPLANTS MAY BEND, BREAK OR LOOSEN DESPITE RESTRICTION IN ACTIVITY. THE PATIENT SHOULD BE ADVISED TO AVOID MECHANICAL VIBRATIONS THAT MAY LOOSEN THE DEVICE. THE PATIENT SHOULD BE ADVISED NOT TO SMOKE OR CONSUME ALCOHOL DURING RECOVERY. POSSIBLE ADVERSE EVENTS LIKE OTHER SPINAL SYSTEM IMPLANTS, THE FOLLOWING ADVERSE EVENTS ARE POSSIBLE. THIS LIST IS NOT EXHAUSTIVE: DELAYED UNION OR NONUNION (PSEUDARTHROSIS). BENDING, DISASSEMBLY OR FRACTURE OF IMPLANT AND COMPONENTS. LOOSENING OF SPINAL FIXATION IMPLANTS MAY OCCUR DUE TO INADEQUATE INITIAL FIXATION, LATENT INFECTION, AND/OR PREMATURE LOADING, POSSIBLY RESULTING IN BONE EROSION, MIGRATION OR PAIN. PAIN, DISCOMFORT, OR ABNORMAL SENSATIONS DUE TO THE PRESENCE OF THE DEVICE. PRESSURE ON SKIN WHERE INADEQUATE TISSUE COVERAGE EXISTS OVER THE IMPLANT, WITH POTENTIAL EXTRUSION THROUGH THE SKIN. DURAL LEAK REQUIRING SURGICAL REPAIR. CESSATION OF GROWTH OF THE FUSED PORTION OF THE SPINE. SUBSIDENCE OF THE IMPLANT INTO ADJACENT BONE. LOSS OF PROPER SPINAL CURVATURE, CORRECTION, HEIGHT AND/OR REDUCTION. INCREASED BIOMECHANICAL STRESS ON ADJACENT LEVELS. IMPROPER SURGICAL PLACEMENT OF THE IMPLANT CAUSING STRESS SHIELDING OF THE GRAFT OR FUSION MASS. INTRAOPERATIVE FISSURE, FRACTURE, OR PERFORATION OF THE SPINE. POSTOPERATIVE FRACTURE DUE TO TRAUMA, DEFECTS, OR POOR BONE STOCK. SERIOUS COMPLICATIONS ASSOCIATED WITH ANY SURGERY MAY OCCUR. THESE INCLUDE, BUT ARE NOT LIMITED TO: WOUND COMPLICATIONS, INFECTION, GENITOURINARY DISORDERS, GASTROINTESTINAL DISORDERS, VASCULAR DISORDERS, INCLUDING THROMBUS; BRONCHOPULMONARY DISORDERS, INCLUDING EMBOLI; BURSITIS, HEMORRHAGE, MYOCARDIAL INFARCTION, PARALYSIS OR DEATH.

Description of Event or Problem · 1

THE PATIENT UNDERWENT SPINAL FIXATION SURGERY ON (B)(6) 2017 WHICH CONSISTED OF POLYAXIAL SCREWS, AXIAL ROD TO ROD CONNECTORS, LOCKING CAPS, AND RODS FROM THE CORAL SPINAL SYSTEM, LEVELS T10-PELVIS. ON (B)(6) 2020, SEASPINE WAS MADE AWARE OF A REVISION SURGERY THAT OCCURRED ON (B)(6) 2017 AS A RESULT OF POSTOPERATIVE PROXIMAL JUNCTIONAL FAILURE WITH A SCREW MIGRATION AT PREVIOUSLY FUSED CONSTRUCT LEVEL T9-T10.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
680501 CORAL SPINAL SYSTEM POLYAXIAL SCREW NKB SEASPINE INC.

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention