BIPOLAR LEAD
Report
- Report Number
- 1644487-2008-00809
- Event Type
- Malfunction
- Date Received
- March 28, 2008
- Date of Event
- February 1, 2008
- Report Date
- February 27, 2008
- Manufacturer
- CYBERONICS, INC.
- Product Code
- LYJ
- PMA / PMN Number
- P970003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
MANUFACTURER REVIEWED X-RAYS OF IMPLANTED DEVICE. RESULT: X-RAYS REVIEWED BY THE MANUFACTURER, NO GROSS LEAD DISCONTINUITIES VISUALIZED. CONCLUSION: DEVICE FAILURE IS SUSPECTED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.
REPORTER INDICATED THAT ON PATIENT'S RECENT OFFICE VISIT, SYSTEM AND NORMAL MODE DIAGNOSTICS HAD RESULTED IN HIGH IMPEDANCE WITH DCDC=7. NO ADVERSE EVENTS WERE REPORTED BY THE PT. ADDITIONALLY, X-RAYS WERE REVIEWED BY THE MANUFACTURER. THREE TIE-DOWNS WERE VISUALIZED AND THE PLACEMENT OF THE GENERATOR WAS NORMAL. THE CONNECTOR PIN(S) DID NOT APPEAR TO BE FULLY INSERTED IN THE CONNECTOR BLOCK, WHICH MAY BE A POTENTIAL CAUSE FOR THE HIGH IMPEDANCE RESULTS. THE FILTER FEEDTHROUGHS APPEARED INTACT. THE STRAIN RELIEF LOOP WAS VISUALIZED; HOWEVER, THE MANUFACTURER WAS UNABLE TO ASSESS WHETHER THE STRAIN RELIEF BEND WAS PRESENT DUE TO THE QUALITY OF THE X-RAYS. A PORTION OF THE LEAD WAS BEHIND THE GENERATOR, AND THEREFORE THIS PORTION COULD NOT BE ASSESSED. NO LEAD DISCONTINUITIES AND ACUTE ANGLES WERE IDENTIFIED ON THE PORTIONS OF THE LEAD THAT WERE VISUALIZED. HOWEVER, THE ELECTRODE REGION AND STRAIN RELIEF BEND COULD NOT BE ASSESSED DUE TO THE CONTRAST OF THE X-RAY. IN ADDITION, IT WAS REPORTED THAT THE PT HAD FALLEN AND WAS INVOLVED IN A MINOR CAR ACCIDENT JUST PRIOR TO THE DISCOVERY OF THE HIGH IMPEDANCE RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BIPOLAR LEAD | NONE | LYJ | CYBERONICS, INC. | 302-30 | 6267 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 21 YR |