FDA Adverse Event Malfunction Summary report: N

TRUE METRIX

MDR report key: 10219901 · Received July 1, 2020

Report

Report Number
1000113657-2020-00449
Event Type
Malfunction
Date Received
July 1, 2020
Date of Event
June 4, 2020
Report Date
August 5, 2020
Manufacturer
TRIVIDIA HEALTH INC
Product Code
NBW
UDI-DI
00021292007836
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

SECTIONS WITH ADDITIONAL INFORMATION AS OF 05-AUG-2020: H10: METER WAS RETURNED FOR EVALUATION; NO DEFECT WAS DETECTED. TEST STRIPS WERE NOT RETURNED FOR EVALUATION.

Additional Manufacturer Narrative · 1

INTERNAL REPORT REFERENCE NUMBER: (B)(4). PRODUCTS WERE RETURNED AND PENDING QC EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-58: USER HAD AN INACCURATE REFERENCE: SELF: THE PERSON IS USING THEMSELVES AS THE REFERENCE OR HOW THEY FEEL AT THE TIME THEY RUN THE BLOOD TEST. NOTE: MANUFACTURER CONTACTED CUSTOMER IN A FOLLOW-UP CALL ON 16-JUN-2020 TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - ABLE TO ESTABLISH CONTACT WITH CUSTOMER AND STATED REPLACEMENT PRODUCT RESOLVED INITIAL CONCERN.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS OBTAINED OF 219, 222, 200, 168 AND 210 MG/DL. THE CUSTOMER¿S EXPECTED AM FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 108-130 MG/DL AND PM EXPECTED BLOOD GLUCOSE TEST RESULT RANGE IS 160-165 MG/DL. THE CUSTOMER WAS NOT USING PROPER TESTING TECHNIQUE. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER FASTING AND PRODUCED TEST RESULTS OF 219 MG/DL AND 235 MG/DL USING TRUEMETRIX METER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 11/26/2020 AND OPEN VIAL DATE IS 05/21/2020. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
684800 TRUE METRIX SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER NBW TRIVIDIA HEALTH INC STRIP, SUNMARK TMX 100CTMG/DL MW3333S 00021292007836

Patients

Seq Age Sex Outcome Treatment
1