OLYMPUS
Report
- Report Number
- 8010047-2008-00055
- Event Type
- Malfunction
- Date Received
- March 28, 2008
- Report Date
- February 28, 2008
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORPORATION
- Product Code
- FTJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE REFERENCED IN THIS REPORT WAS RETURNED TO OLYMPUS FOR INVESTIGATION. THE INVESTIGATION CONFIRMED THE USER'S REPORT OF IMAGE DIFFICULTY. THE IMAGE WAS FOUND TO CONTAIN STATIC AND FLICKER INTERMITTENTLY AFTER THE DEVICE WAS TESTED FOR SEVERAL HOURS. THERE WAS EVIDENCE OF FLUID FOUND INSIDE THE ENDOSCOPE CONNECTOR, AND A CRACK NOTED ON THE LIGHT GUIDE LENS. A LARGE DENT WAS OBSERVED ON THE DISTAL END COVER. ADDITIONALLY, THE VARIABLE STIFFNESS AND ANGULATION CONTROL KNOBS WERE DETERMINED TO NOT FUNCTION PROPERLY DUE TO STRETCHED COIL AND ANGLE WIRES, LIKELY DUE TO EXTENDED USE. THE BENDING COVER WAS DISASSEMBLED AND BROKEN/FRAYED WIRES WERE FOUND ON THE BENDING SECTION MESH, ALSO DUE TO EXTENSIVE USAGE. FOLLOWING AERATION, THE UNIT WAS RETESTED AND THE IMAGE WAS FOUND TO BE WITHIN STANDARD. AS THE UNIT PASSED LEAK TESTING, USER HANDLING CANNOT BE RULED OUT AS THE SOURCE OF THE FLUID INVASION. THIS REPORT IS BEING FILED AS AN MDR IN AN ABUNDANCE OF CAUTION.
THE USER FACILITY REPORTED THAT DURING A COLONOSCOPY, A COMPLETE LOSS OF IMAGE WAS EXPERIENCED, AND THE IMAGE COULD NOT BE RECOVERED. THE PROCEDURE WAS REPORTEDLY COMPLETED WITH A DIFFERENT, BUT SIMILAR DEVICE. THERE WAS NO PATIENT INJURY AND NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OLYMPUS | EVIS EXERA CLONOVIDEOSCOPE | FTJ | OLYMPUS MEDICAL SYSTEMS CORPORATION | CF-Q160AL | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |