FDA Adverse Event Malfunction Summary report: N

LIFEPAK 500 AUTOMATED EXTERNAL DEFIBRILLATOR

MDR report key: 1021911 · Received March 28, 2008

Report

Report Number
3015876-2008-00239
Event Type
Malfunction
Date Received
March 28, 2008
Date of Event
February 27, 2008
Report Date
March 3, 2008
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
MKJ
PMA / PMN Number
K955854
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL EVALUATED THE DEVICE, VERIFIED THE REPORTED INTERMITTENT FAILURE, AND OBSERVED INTERMITTENT OPERATION OF THE ON/OFF BUTTON. THE SWITCH WAS REPAIRED AND THEN PROPER DEVICE OPERATION WAS OBSERVED THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WOULD NOT POWER ON. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK 500 AUTOMATED EXTERNAL DEFIBRILLATOR MKJ PHYSIO-CONTROL, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA