FDA Adverse Event Injury Summary report: N

GMK-SPHERE TIBIAL TRAY FIXED CEMENTED SIZE 5 R

MDR report key: 10217918 · Received July 1, 2020

Report

Report Number
3005180920-2020-00369
Event Type
Injury
Date Received
July 1, 2020
Date of Event
June 2, 2020
Report Date
July 1, 2020
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JWH
UDI-DI
07630030819933
PMA / PMN Number
K090988
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED BY MEDACTA REGULATORY AFFAIRS DEPARTMENT ON 12 JUNE 2020: LOT 1900539: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 28-MAY-2019. EXPIRATION DATE: 2024-05-18. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITH NO OTHER SIMILAR REPORTED EVENT. ADDITIONAL IMPLANT INVOLVED: GMK-SPHERE 02.12.0512FR TIBIAL INSERT FIXED SPHERE FLEX SIZE 5/12 MM R (K121416) LOT. 1811504. BATCH REVIEW PERFORMED BY MEDACTA REGULATORY AFFAIRS DEPARTMENT ON 12 JUNE 2020: LOT 1811504: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 28-MAR-2019. EXPIRATION DATE: 2024-03-10. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITH NO OTHER SIMILAR REPORTED EVENT. ADDITIONAL IMPLANT INVOLVED: GMK-SPHERE 02.12.0025R FEMORAL COMPONENT SPHERE CEMENTED SIZE 5+ R (K140826) LOT. 1900333. BATCH REVIEW PERFORMED BY MEDACTA REGULATORY AFFAIRS DEPARTMENT ON 12 JUNE 2020: LOT 1900333: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 23-APR-2019. EXPIRATION DATE: 2024-04-08. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITH NO OTHER SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

THE PATIENT CAME IN REPORTING PAIN 10 MONTHS AFTER THE PRIMARY SURGERY POSSIBLY DUE TO AN OVERSIZED TIBIA OR UNDERSIZED FEMUR. THE SURGEON REVISED THE SIZE 5 TIBIAL TRAY WITH A SIZE 4 TIBIAL TRAY, THE SIZE 5+ FEMORAL COMPONENT WITH A SIZE 5 FEMORAL COMPONENT AND SIZE 5/12MM INSERT WITH A SIZE 4/20MM INSERT. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
683664 GMK-SPHERE TIBIAL TRAY FIXED CEMENTED SIZE 5 R FIXED TIBIAL TRAY JWH MEDACTA INTERNATIONAL SA 02.07.1205R 1900539 07630030819933

Patients

Seq Age Sex Outcome Treatment
1 48 YR Required Intervention