FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 1021747 · Received March 19, 2008

Report

Report Number
1720753-2008-19775
Event Type
Malfunction
Date Received
March 19, 2008
Date of Event
March 6, 2008
Report Date
March 18, 2008
Manufacturer
GE OEC MEDICAL SYSTEMS, INC.
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A GE SVC REP PERFORMED AN ON SITE INVESTIGATION. THE MALFUNCTION WAS VERIFIED. REPLACED THE DISPLAY ADAPTER PCB AND PLACED THE GIGABIT CABLE IN THE CORRECT POSITION ON THE RTOS PCB. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PLACED BACK INTO SVC.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE 9900 SYSTEM WILL NOT BOOT. THERE WAS NO REPORT OF PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS, INC. 9900 NA

Patients

Seq Age Sex Outcome Treatment
1